5
Service Information
5-1
WARNING
The information in this section is in-
tended only for quali
Þ
ed service per-
sonnel. Some of the procedures may
expose you to hazardous voltages
that could result in personal injury or
death. Do not attempt to perform
these procedures unless you are
quali
Þ
ed to do so.
5.1
Introduction
This section contains information necessary to service
the Model 7012 matrix card and is arranged as follows:
5.2
Handling and cleaning precautions:
Discusses
handling procedures and cleaning methods for
the matrix card.
5.3
Performance veri
Þ
cation:
Covers the procedures
necessary to determine if the card is operating
properly.
5.4
Special handling of static-sensitive devices:
Reviews precautions necessary when handling
static-sensitive devices.
5.5
Principles of operation:
Brie
ß
y discusses circuit
operation.
5.6
Troubleshooting:
Presents some troubleshooting
tips for the matrix card.
5.2
Handling and cleaning precautions
Because of the high impedance circuits on the Model
7012, care should be taken when handling or servicing
the card to prevent possible contamination, which
could degrade performance. The following precau-
tions should be taken when handling the matrix card.
Do not store or operate the card in an environment
where dust could settle on the circuit board. Use dry ni-
trogen gas to clean dust off the card if necessary.
Handle the card only by the side edges. Do not touch
any board surfaces, components, or connectors. Do not
touch areas adjacent to electrical contacts. When servic-
ing the card, wear clean cotton gloves.
If making solder repairs on the circuit board, use an
OA-based (organic activated)
ß
ux. Remove the
ß
ux
from these areas when the repair is complete. Use pure
water along with plenty of clean cotton swabs to re-
move the
ß
ux. Take care not to spread the
ß
ux to other
areas of the circuit board. Once the
ß
ux has been re-
moved, swab only the repaired area with methanol,
then blow dry the board with dry nitrogen gas.
After cleaning, the card should be placed in a 50
°
C low
humidity environment for several hours.
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com
Содержание 7012-S
Страница 62: ...Operation 4 16 Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Страница 83: ...2 25917...
Страница 84: ...TC17 100 25917 2...
Страница 86: ......
Страница 87: ......
Страница 89: ......
Страница 90: ......