7-11
Pattern screen 4
A vertical bar moves through the
Þ
ve columns of
the
Þ
rst 5
×
7 character position. One and only one
bar must be visible at a time. Movement will con-
tinue until any key is pressed.
Pattern screen 5
All 35 dots of each 5
×
7 character block will turn on
one character at a time. The upper row blocks in-
clude a number and (except for positions 5, 9, 13,
and 16) an annunciator. Verify that one and only
one block illuminates fully for each character posi-
tion.
The
Þ
rst test pattern automatically appears when
the display test commences. Press any key except
EXIT to move to each successive pattern.
6. Verify all
Þ
ve patterns, then press EXIT to return to
the TEST MENU.
7.6.7 Display test — char set
This test is used to display the ASCII character set.
1. Press the MENU key.
2. Select the TEST option and press ENTER.
3. Under the SELF TEST MENU, select the DISPLAY
TESTS option.
4. Under DISPLAY TESTS, select CHAR-SET and
press ENTER to display the ASCII charatcer set.
5. Press any key (except EXIT) to scroll through the
character set.
Figure 7-1
First pattern for display test
REM
TALK
LSTN
SRQ
ARM
ERR
EDIT
REAR
REL
FILT
MATH
4W
AUTO
TRIG
*
SMPL
6. When
Þ
nished, use the EXIT key to back out of the
menu structure.
7.7
Test descriptions and notes
The following information provides additional details
for interpreting any error messages which may result
during the Built-In-Tests. These notes are intended to
help you isolate faults to the board level, rather than
the component level.
7.7.1 Initial power-up test
Upon power-up, the Model 7001 writes test patterns to
all volatile memory locations and checks that they read
back properly. It then reads all ROM locations and ver-
i
Þ
es that a calculated checksum matches a checksum
stored in ROM. If either of these tests fail, the digital
board processor will continue to loop on these tests. If
any of these tests fail, the display will indicate “No
Comm Link” after the Model 7001 has been on for sev-
eral seconds. This message indicates that the digital
board has not communicated with the display board
within the time-out interval.
After memory tests, the Model 7001 checks con
Þ
gura-
tion data stored in non-volatile memory, con
Þ
gures the
instrument according to power-up setups, and dis-
plays startup information. Con
Þ
guration data is stored
such that the Model 7001 may determine if it is valid.
When a Model 7001 has power applied with uninitial-
ized memory, it may brie
ß
y display a “con
Þ
guration
error” message.
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com
Содержание 7001
Страница 183: ...IEEE 488 Reference 5 64 Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Страница 185: ...IEEE 488 Reference 5 66 Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Страница 239: ...IEEE 488 Reference 5 120 Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Страница 284: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Страница 285: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Страница 286: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Страница 287: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Страница 288: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Страница 293: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Страница 294: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Страница 295: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Страница 296: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Страница 298: ...Replaceable Parts 8 6 Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Страница 299: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Страница 300: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Страница 301: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Страница 303: ...Replaceable Parts 8 8 Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Страница 304: ......
Страница 305: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Страница 307: ...Replaceable Parts 8 10 Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Страница 308: ......
Страница 309: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Страница 311: ...Replaceable Parts 8 12 Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Страница 312: ......
Страница 313: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Страница 315: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...