![Keithley 6485 Скачать руководство пользователя страница 66](http://html2.mh-extra.com/html/keithley/6485/6485_user-manual_4060113066.webp)
3-10
Measurements and Sourcing Voltage
Model 6485 and 6487 User’s Manual
Step 2. Perform zero correction
To achieve optimum accuracy for high resistance measurements, it is recommended that
you zero correct the picoammeter before enabling the ohms function. To do so, make sure
that zero check and the 2nA range are selected, then press the REL key to perform zero
correction (MON indicator on).
Step 3. Select a manual current range or enable auto range
Use the manual RANGE keys to select a manual current measurement range or press
AUTO to enable auto range. When using manual ranging, choose an appropriate value
based on the voltage source setting and the expected measured resistance (I = V/R).
Step 4. Connect the resistance to be measured to the picoammeter
Basic connections for ohms measurements are shown in
. Note that both the
picoammeter INPUT and the V-SOURCE OUTPUT jacks are connected to the resistance
under test.
WARNING
A safety shield is advisable whenever measurements are being made
with voltages over 30V DC. Connections for the safety shield are
shown in
. The metal safety shield must completely sur-
round the noise shield or floating test circuit, and it must be connected
to safety earth ground using #18 AWG or larger wire.
Step 5. Select ohms function
Press the I|
Ω
key to make sure the ohms function is selected.
Step 6. Turn on voltage source
Press the OPER key to turn on the voltage source output. The VOLTAGE SOURCE
OPERATE indicator will turn on.
Step 7. Disable zero check and take a reading from the display
If the readings are noisy, use filtering to reduce noise. It may be advisable to use the
alternating voltage ohms or add shielding.
NOTE
For any ohms measurements, the ohms reading is invalid and unknown if the
voltage source is in compliance. Therefore, a value of -9.9e+36 will be returned
over the GPIB and the message I-LIMIT will be displayed on the front panel for
both normal readings and buffer recall readings for any ohms readings where
the voltage source went into compliance.
Содержание 6485
Страница 40: ...1 28 Getting Started Model 6485 and 6487 User s Manual...
Страница 94: ...7 8 Remote Operation and Commands Model 6485 and 6487 User s Manual...
Страница 95: ...A Specifications...
Страница 98: ...A 4 Specifications Model 6485 and 6487 User s Manual...
Страница 105: ...C ExamplePrograms Programming examples lists and defines three types of programming exam ples...
Страница 113: ......
Страница 114: ......