Model 6485 Picoammeter Instruction Manual
Relative, mX+b, m/X+b (reciprocal), and log
5-5
If the instrument is programmed to perform an infinite number of measurements (arm
count or trigger count set to infinite), you cannot use the :DATA? command to return
Rel’ed readings. However, you can use the :DATA:LATest? command to return the last
Rel’ed reading after aborting the measurement process. After sending the INITiate com-
mand to start the measurement process, use the ABORt command to abort the measure-
ment process, then use :DATA:LATest? to return to the last Rel’ed reading.
Programming example — relative
This program fragment establishes a 1
µ
A baseline for measurements:
CALC2:NULL:OFFS 1e-6
' Set Rel value of 1
µ
A.
CALC2:NULL:STAT ON
' Enable Rel.
CALC2:FEED SENS
' Rel input signal.
INIT
' Trigger reading(s).
CALC2:DATA?
' Request Rel’ed reading.
mX+b, m/X+b (reciprocal), and Logarithmic
mX+b and m/X+b
The following math operations manipulate normal display readings (X) mathematically
according to the following calculations:
Y = mX+b
Y = m/X+b
where: X is the normal display reading
m and b are user-entered constants for scale factor and offset
Y is the displayed result
NOTE
Changing the “m” or “b” for mX+b also changes it for m/X+b.
Pressing the mX + b key toggles the state of the mX+b math calculation on or off. When
pressed, the MATH annunciator will toggle in addition to the presently selected units des-
ignator (default is an “X”) indicating the state of the mX+b math function (note that the m/
X+b calculation uses the same designator as selected for the mX+b calculation).
Configuring and controlling mX+b and m/X+b
To configure and control either of these math calculations, perform the following steps:
NOTE
Enabling or disabling by pressing the mX+b key disables Rel (if Rel is enabled).
1.
Press CONFIG and then MX+B (or M/X+B) to display the present scale factor:
M: +1.0000000 ^ (factory default)
Содержание 6485
Страница 1: ...Model 6485Picoammeter Instruction Manual A G R E A T E R M E A S U R E O F C O N F I D E N C E ...
Страница 20: ......
Страница 24: ......
Страница 44: ...1 20 Getting Started Model 6485 Picoammeter Instruction Manual ...
Страница 66: ...3 6 Measurements Model 6485 Picoammeter Instruction Manual ...
Страница 138: ...9 20 Remote Operation Model 6485 Picoammeter Instruction Manual ...
Страница 159: ...11 CommonCommands ...
Страница 164: ...11 6 Common Commands Model 6485 Picoammeter Instruction Manual ...
Страница 165: ...12 SCPISignalOriented MeasurementCommands ...
Страница 190: ...14 12 SCPI Reference Tables Model 6485 Picoammeter Instruction Manual ...
Страница 214: ...16 14 Calibration Model 6485 Instruction Manual ...
Страница 219: ...A Specifications ...
Страница 221: ...B StatusandErrorMessages ...
Страница 227: ...C GeneralMeasurement Considerations ...
Страница 233: ...D DDCEmulationCommands ...
Страница 247: ...E ExamplePrograms ...
Страница 250: ...E 4 Example Programs Model 6485 Picoammeter Instruction Manual ...
Страница 251: ...F IEEE 488BusOverview ...
Страница 266: ...F 16 IEEE 488 Bus Overview Model 6485 Picoammeter Instruction Manual ...
Страница 267: ...G IEEE 488andSCPI ConformanceInformation ...
Страница 311: ......