DDC Emulation Commands
Model 6485 Picoammeter Instruction Manual
A.
The buffer size command (In) is not used by the Model 485. The buffer size com-
mand is similar to the SCPI TRACe:POINts command (see
B.
The moving filter cannot be selected from the DDC language.
C.
For Model 6485, the terminator commands (Y) to set the terminator are different
from the Y commands used by the 485. Also note that the “YY” response to the U0
command is different.
D.
For the Model 6485, the Buffer Full bit in the U1X status word does not get cleared
until either the buffer is resized or buffer storage is reactivated. Note that requesting
a buffer reading does not clear the U1 Buffer Full bit.
E.
The U2 status command is not used by the Model 485. The response message indi-
cates the buffer size (In) and the actual number of readings stored in the buffer.
F.
Calibration must be already unlocked and enabled.
G.
See CALibration subsystem.
Status
Word
U0
Return status word. See
must be queried individually. For example:
SENSe[1]:CURRent:RANGe?
U1
NEW
U2
Buffer size and readings stored. See
D, NEW
Digital
Calibration
V
Calibration commands (V+n.nnnnE+nn) not supported G
Execute
X
Execute other device-dependent commands. SCPI not
applicable (SCPI commands execute as they are
received)
Terminator
Y(ASCII)
Not Available
C, NDS
Y0
LFCR (line feed, carriage return)
C, NDS
Y1
CRLF (carriage return, line feed)
C, NDS
X
Y2
LF (line feed)
C, NDS
Y3
CR (carriage return)
C, NDS
Y4
None
C, NDS
Baseline
Suppres-
sion
(Rel)
Z0
Suppression (Rel) disabled
CALCulate2:NULL:STATe OFF
Z1
Suppression (Rel) enabled
CALCulate2:NULL:ACQuire
CALCulate2:NULL:STATe ON
Table D-1 (continued)
Device dependent command summary
Mode
Command
Description
Equivalent SCPI commands
Note
Default
Содержание 6485
Страница 1: ...Model 6485Picoammeter Instruction Manual A G R E A T E R M E A S U R E O F C O N F I D E N C E ...
Страница 20: ......
Страница 24: ......
Страница 44: ...1 20 Getting Started Model 6485 Picoammeter Instruction Manual ...
Страница 66: ...3 6 Measurements Model 6485 Picoammeter Instruction Manual ...
Страница 138: ...9 20 Remote Operation Model 6485 Picoammeter Instruction Manual ...
Страница 159: ...11 CommonCommands ...
Страница 164: ...11 6 Common Commands Model 6485 Picoammeter Instruction Manual ...
Страница 165: ...12 SCPISignalOriented MeasurementCommands ...
Страница 190: ...14 12 SCPI Reference Tables Model 6485 Picoammeter Instruction Manual ...
Страница 214: ...16 14 Calibration Model 6485 Instruction Manual ...
Страница 219: ...A Specifications ...
Страница 221: ...B StatusandErrorMessages ...
Страница 227: ...C GeneralMeasurement Considerations ...
Страница 233: ...D DDCEmulationCommands ...
Страница 247: ...E ExamplePrograms ...
Страница 250: ...E 4 Example Programs Model 6485 Picoammeter Instruction Manual ...
Страница 251: ...F IEEE 488BusOverview ...
Страница 266: ...F 16 IEEE 488 Bus Overview Model 6485 Picoammeter Instruction Manual ...
Страница 267: ...G IEEE 488andSCPI ConformanceInformation ...
Страница 311: ......