Section 17: Calibration
Model 6482 Dual-Channel Picoammeter / Voltage Source Reference Manual
Performance verification
WARNING
The information in this section is intended only for qualified service
personnel. Some of the procedures may expose you to hazardous voltages
that could result in personal injury or death. Do not attempt to perform these
procedures unless you are qualified to do so.
Use the procedures in this section to verify that the Keithley Instruments Model 6482’s performs
within the limits stated in the instrument’s one-year accuracy specifications. Perform the
verification procedures:
•
When you first receive the instrument to make sure that it was not damaged during
shipment.
•
To verify that the instrument meets factory specifications.
•
To determine if calibration is required.
•
After performing a calibration adjustment to make sure the instrument was adjusted
properly.
NOTE
If the instrument is still under warranty and its performance is outside specified limits,
contact your Keithley Instruments representative or the factory to determine the correct
course of action.
Verification test requirements
Be sure to perform the verification tests:
•
Under the proper environmental conditions.
•
After the specified warm-up period.
•
Using the correct line voltage.
•
Using the proper test equipment.
•
Using the specified output signals and reading limits.
NOTE
Product specifications are subject to change. Listed uncertainties and test limits are
provided only as examples. Always verify values against actual product specifications.
Environmental conditions
Conduct performance verification procedures in a test environment with:
•
An ambient temperature of 18 °C to 28 °C (65 °F to 82 °F).
•
A relative humidity of less than 70 percent, unless otherwise noted.
NOTE
Product specifications that are listed as 18 ºC to 28 ºC assume adjustment has been
done at 23 ºC. If the Model 6482 is adjusted at a different temperature, the specifications
apply to ±5 ºC of that adjustment temperature.