Section 16: SCPI Command Reference
Configure and control buffer
FREE?
:TRACe:FREE?
Read status of memory
Description
This command is used to read the status of storage memory. After sending this
command and addressing the Model 6482 to talk, two values separated by
commas are sent to the computer. The first value indicates how many bytes of
memory are available, and the second value indicates how many bytes are
reserved to store readings.
POINts <n>
:TRACe:POINts <n>
Specify buffer size
Parameters
<n> =
1 to 3000
Specify buffer size
MINimum
1
MAXimum
3000
DEFault
100
Query
:POINts?
Query buffer size
:POINts? MINimum
Query smallest allowable buffer size
:POINts? MAXimum
Query largest allowable buffer size
:POINts? DEFault
Query *RST default buffer size
Description
This command is used to specify the size of the buffer.
ACTual?
:TRACe:POINts:ACTual?
Query number of stored readings
Description
This query command is used to determine how many stored readings are in the
buffer. After sending this command and addressing the instrument to talk, the
number of readings stored in the buffer will be sent to the computer.
CONTrol <name>
:TRACe:FEED:CONTrol <name>Start or stop buffer
Parameters
<name> = NEXT
Fills buffer and stops
NEVer
Disables buffer storage
Query
:CONTrol?
Query buffer control
Description
This command is used to select the buffer control. When NEXT is selected, the
asterisk (*) indicator turns on to indicate that the buffer is enabled. The storage
process starts when Model 6482 is taken out of idle to perform source-measure
operations.
After the buffer stores the specified number of reading arrays (as set by the
:POINts command), the asterisk annuciator turns off to indicate that storage is
done.
With NEVer selected, storage into the buffer is disabled.