SOURce3 <name>
:FORMat:SOURce3 <name>
Set SOUR3 and TTL response formats
Parameters
<name> = ASCii
ASCII format
HEXadecimal
Hexadecimal format
OCTal
Octal format
BINary
Binary format
Query
:SOURce3?
Query response format
Description
This command controls the response format for all CALC7:XXXX:SOUR3 and
SOUR3:TTL queries in a manner similar to formats set by the FORM:SREG
command. See
and
“SOURce subsystem,” page Section
topics for details.
CALC data elements
CALCulate <item list>
:FORMat:ELEMents:CALCulate <item list>Set CALC data elements
Parameters
<item list> =CALCulate
Include CALC1 to CALC7 data
TIME
Include timestamp
STATus
Include status information
NOTE
Each item in the list must be separated by a comma (for example, :CALCulate
CALC,TIME,STAT).
Query
:CALCulate?
Query CALC data element list
Description
This command allows you specify the data elements returned with the
CALC1:DATA? through CALC7:DATA? queries. With CALC specified, the
corresponding CALC1 through CALC7 element will be included in the response.
With TIMe or STATus specified, timestamp and status information will be returned
with the queries. See
“Calculate subsystems,” page Section 16-16,
for a complete
description of the various CALC1 through CALC7 elements. See also
elements,” page Section 16-29,
for a description of TIME and STATus information.
TRACe data elements
TRACe <item list>
:FORMat:ELEMents:TRAce <item list>Specify data elements for buffer
Parameters
<item list> =CURRent[1]
Includes channel 1 current reading
CURRent2
Includes channel 2 current reading
CALCulate1
Includes CALC1 element
CALCulate2
Includes CALC2 element
CALCulate3
Includes CALC3 element
CALCulate4
Includes CALC4 element
CALCulate5
Includes CALC5 element
CALCulate6
Includes CALC6 element
CALCulate7
Includes CALC7 element
TIME
Includes timestamp
STATus
Includes status information
ALL
Includes all data elements
DEFault
Includes only CURR1 and CURR2
NOTE
Each item in the list must be separated by a comma (for example, CURR1, CALC2,
TIME).