Configure and control limit tests
COMPliance:FAIL <name>
:CALCulate7:LIMit[1]:COMPliance:FAIL <name>Set channel 1 compliance fail
mode
:CALCulate7:LIMit2:COMPliance:FAIL <name>Set channel 2 compliance fail
mode
Parameters
<name> = IN
Fail Limit 1/2 test when instrument goes into
compliance
OUT
Fail Limit 1/2 test when instrument comes out of
compliance
Query
:FAIL?
Query when Limit 1/2 test failure occurs
Description
These commands are used to specify the condition that will cause the Limit 1 and
Limit 2 tests to fail. With IN specified, the test will fail when the Model 6482 goes
into compliance (fixed 20 mA current limit for each voltage source). With OUT
specified, the test will fail when the Model 6482 comes out of compliance.
[:DATA] <n>
:CALCulate7:LIMitx:LOWer[:DATA] <n>Specify lower Limit x (x = 3-6)
:CALCulate7:LIMitx:UPPer[:DATA] <n>Specify upper Limit x (x = 3-6)
Parameters
<n> =
-9.999999e20 to Specify limit value
9.999999e20
DEFault
Set specified lower limit to -1
Set specified upper limit to 1
MINimum
Set specified limit to -9.999999e20
MAXimum
Set specified limit to +9.999999e20
Query
:UPPer?
Query specified upper limit
:UPPer? DEFault
Query *RST default upper limit
:UPPer? MINimum
Query lowest allowable upper limit
:UPPer? MAXimum
Query largest allowable upper limit
:LOWer?
Query specified lower limit
:LOWer? DEFault
Query *RST default lower limit
:LOWer? MINimum
Query lowest allowable lower limit
:LOWer? MAXimum
Query largest allowable lower limit
Description
These commands are used to set the upper and lower limits for the Limit 3
through Limit 6 tests (software limits).
STATe <b>
:CALCulate7:LIMit[1]:STATe <b>Control Limit 1 (channel 1) test
:CALCulate7:LIMit2:STATe <b> Control Limit 2 (channel 2) test
:CALCulate7:LIMitx:STATe <b> Control Limit x test (x = 3-6)
Parameters
<b> =
1 or ON
Enable specified limit test
0 or OFF
Disable specified limit test
Query
:STATe?
Query state of specified limit test
Description
These commands are used to enable or disable Limit 1 through Limit 6, tests.
Limit 1 and Limit 2 are the compliance limit tests for channel 1 and channel 2
respectively, while Limit 3 through Limit 6 are the software limit tests. Any limit test
not enabled is simply not performed.
Содержание 6482
Страница 3: ......
Страница 62: ...Section 4 Range Digits Speed and Filters In this section Topic Page Range and digits 4 2 Speed 4 4 Filters 4 6...
Страница 90: ...Section 7 Sweep Operation In this section Topic Page Sweep types 7 2 Configuring and running a sweep 7 5...
Страница 182: ...Section 14 Common Commands In this section Topic Page Command summary 14 2 Command reference 14 2...