9-6
Return to
6482-901-01 Rev. A / August 2012
Section 9: Measurement Concepts
Model 6482 Dual-Channel Picoammeter / Voltage Source Reference Manual
Figure 9-5
Loading effects
Data flow
Data flow for front-panel operation is summarized by the block diagrams provided in
,
and
shows data flow through the various CALC blocks. Note that if REL is enabled, the
result of the rel operation is sent to the other blocks.
NOTE
See
for remote operation data flow information that describes the CALC
blocks in detail.
A. Normal bias source operation
B. Bias source in compliance
V
S
Operating
point
Voltage source
load line
20 mA
10 V
= 8 V
8 V
400
Ω
DUT lo
ad
lin
e (R)
Output
current (I)
Output voltage (V)
Operating
point
Voltage source
load line
Output voltage (V)
Output
current (I)
10 mA
20 mA
10 V
1 k
Ω
DUT load line (R)
I = V / R
= 10 mA
= (20 mA) (400
Ω
)
= I
M
•
R
= 10 V/1 k
Ω
Current limit
load line
Current limit
load line