Performance Verification
18-29
2.
Connect the Model 6430 ohms function for the 2-wire sense and con
fi
gure guard mode
as follows:
•
Press CONFIG then MEAS
Ω
. The instrument will display the following:
CONFIG OHMS
SOURCE GUARD
•
Select GUARD, and then press ENTER. The following will be displayed:
GUARD
OHMS CABLE
•
Select CABLE, and then press ENTER.
•
Press EXIT to return to normal display.
WARNING
Hazardous voltage may be present on SHIELD when using guarding.
3.
Press MEAS
Ω
to select the ohms measurement function, and make sure the source out-
put is turned on.
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
MADE IN
U.S.A.
INPUT/
OUTPUT
42V
PEAK
250V
PEAK
TRIGGER
LINK
4-WIRE
SENSE
HI
LO
LINE RATING
100-240VAC
50, 60, HZ
100VA MAX
RS232
IEEE-488
(ENTER IEEE ADDRESS
WITH FRONT PANEL MENU)
250V
PEAK
5V
PEAK
5V
PEAK
5V
PK
V,
Ω
,
GUARD
GUARD
SENSE
LINE FUSE
SLOWBLOW
2.5A, 250V
INTERLOCK-
DIGITAL I/O
REMOTE
PreAmp
!
Connect PreAmp
IN/OUT HIGH Jack
Directly to 5156 OUTPUT Jack
BNC Shorting Plug
(Connect to Resistor
Being used)
Note
: Remove Link Between
SHIELD and CHASSIS
Mainframe/PreAmp Cable
Model 6430
100G
Ω
1 nF
100nF
10G
Ω
1G
Ω
100M
Ω
Model 5156 Electrometer Calibration Standard
Figure 18-9
Connections for
Remote PreAmp
2G
Ω
-200G
Ω
range verification
Содержание 6430
Страница 26: ......
Страница 32: ......
Страница 78: ...2 14 Connections ...
Страница 98: ...3 20 Basic Source Measure Operation ...
Страница 138: ...5 30 Source Measure Concepts ...
Страница 156: ...6 18 Range Digits Speed and Filters ...
Страница 168: ...7 12 Relative and Math ...
Страница 176: ...8 8 Data Store ...
Страница 202: ...9 26 Sweep Operation ...
Страница 248: ...11 22 Limit Testing ...
Страница 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Страница 418: ...17 108 SCPI Command Reference ...
Страница 450: ...18 32 Performance Verification ...
Страница 477: ...A Specifications ...
Страница 489: ...B StatusandErrorMessages ...
Страница 498: ...B 10 Status and Error Messages ...
Страница 499: ...C DataFlow ...
Страница 503: ...D IEEE 488BusOverview ...
Страница 518: ...D 16 IEEE 488 Bus Overview ...
Страница 519: ...E IEEE 488andSCPI ConformanceInformation ...
Страница 523: ...F MeasurementConsiderations ...
Страница 539: ...G GPIB488 1Protocol ...
Страница 557: ......