17-60
SCPI Command Reference
Set measurement speed
:NPLCycles <n>
[:SENSe[1]]:CURRent[:DC]:NPLCycles <n>
Set speed (PLC)
[:SENSe[1]]:VOLTage[:DC]:NPLCycles <n>
Set speed (PLC)
[:SENSe[1]]:RESistance:NPLCycles <n>
Set speed (PLC)
Parameters
<n> =
0.01 to 10
Power-line cycles per integration
DEFault
10
MINimum
0.01
MAXimum
10
Query
:NPLCycles?
Query programmed PLC value
:NPLCycles? DEFault
Query *RST default PLC
:NPLCycles? MINimum
Query minimum PLC
:NPLCycles? MAXimum
Query maximum PLC
Description
This command is used to set the integration period (speed) for measure-
ments. NPLC (Number of Power Line Cycles) expresses the integration
period by basing it on the power line frequency. For example, for a PLC of
1, the integration period would be 1/60 (for 60Hz line power) which is 16.67
msec.
Note that this is a global command. Thus, if you set the speed for voltage
measurements to 10 PLC, then current and resistance will also set to 10
PLC.
Con
fi
gure and control
fi
lters
NOTE
Detailed information on the repeat, median and moving filters are provided in Sec-
tion 6,“Filters.”
:AUTO <b>
[:SENSe[1]]:AVERage:AUTO <b>
Enable/disable auto
fi
lter
Parameters
<b> =
0 or OFF
Disable auto
fi
lter
1 or ON
Enable auto
fi
lter
Query
:AUTO?
Query state of auto
fi
lter
Description
With auto
fi
lter enabled, the instrument automatically selects
fi
lter settings
for current measurements. Heavy
fi
ltering is used for the low current ranges,
and less
fi
ltering as the current range increases. The auto
fi
lter settings are
listed in Tables 6-5, 6-6, and 6-7. For voltage measurements, auto
fi
lter sets
the repeat and moving count to one, and the median rank to zero. These set-
tings effectively disable the three
fi
lters.
When auto
fi
lter is disabled, the present count and rank settings for the three
fi
lters are used for all measurement functions and ranges.
Содержание 6430
Страница 26: ......
Страница 32: ......
Страница 78: ...2 14 Connections ...
Страница 98: ...3 20 Basic Source Measure Operation ...
Страница 138: ...5 30 Source Measure Concepts ...
Страница 156: ...6 18 Range Digits Speed and Filters ...
Страница 168: ...7 12 Relative and Math ...
Страница 176: ...8 8 Data Store ...
Страница 202: ...9 26 Sweep Operation ...
Страница 248: ...11 22 Limit Testing ...
Страница 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Страница 418: ...17 108 SCPI Command Reference ...
Страница 450: ...18 32 Performance Verification ...
Страница 477: ...A Specifications ...
Страница 489: ...B StatusandErrorMessages ...
Страница 498: ...B 10 Status and Error Messages ...
Страница 499: ...C DataFlow ...
Страница 503: ...D IEEE 488BusOverview ...
Страница 518: ...D 16 IEEE 488 Bus Overview ...
Страница 519: ...E IEEE 488andSCPI ConformanceInformation ...
Страница 523: ...F MeasurementConsiderations ...
Страница 539: ...G GPIB488 1Protocol ...
Страница 557: ......