Relative and Math
7-9
Table 7-4
Voltage coefficient programming example
Command
Description
*RST
:SENS:FUNC:ON:ALL
:SENS:RES:MODE MAN
:SOUR:FUNC VOLT
:SOUR:VOLT:STAR 10
:SOUR:VOLT:STOP 50
:SOUR:VOLT:MODE SWE
:SOUR:SWE:POIN 2
:TRIG:COUN 2
:CALC:MATH:NAME “VOLTCOEF”
:CALC:STAT ON
:OUTP ON
:INIT
:CALC:DATA?
Reset unit to GPIB defaults.
Enable all sense functions.
Manual resistance mode.
Volts source function.
10V start voltage.
50V stop voltage.
Volts sweep mode.
Sweep points = 2.
Trigger count = 2.
Select voltage coefficient math expression.
Enable math.
Turn on output.
Trigger sweep.
Request voltage coefficient data.
Triax Cable
HI
LO
KEITHLEY
6430
REMOTE
PreAmp
SENSE
MAINFRAME
Connect to REMOTE
PreAmp connector on rear
panel of mainframe
Resistor
Under
Test
Preamp Cable
IN/OUT
HIGH
Figure 7-2
Connections for
voltage coefficient
tests
Содержание 6430
Страница 26: ......
Страница 32: ......
Страница 78: ...2 14 Connections ...
Страница 98: ...3 20 Basic Source Measure Operation ...
Страница 138: ...5 30 Source Measure Concepts ...
Страница 156: ...6 18 Range Digits Speed and Filters ...
Страница 168: ...7 12 Relative and Math ...
Страница 176: ...8 8 Data Store ...
Страница 202: ...9 26 Sweep Operation ...
Страница 248: ...11 22 Limit Testing ...
Страница 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Страница 418: ...17 108 SCPI Command Reference ...
Страница 450: ...18 32 Performance Verification ...
Страница 477: ...A Specifications ...
Страница 489: ...B StatusandErrorMessages ...
Страница 498: ...B 10 Status and Error Messages ...
Страница 499: ...C DataFlow ...
Страница 503: ...D IEEE 488BusOverview ...
Страница 518: ...D 16 IEEE 488 Bus Overview ...
Страница 519: ...E IEEE 488andSCPI ConformanceInformation ...
Страница 523: ...F MeasurementConsiderations ...
Страница 539: ...G GPIB488 1Protocol ...
Страница 557: ......