Model 6220/6221 Reference Manual
Sweeps
4-11
Return to
Performing a linear staircase sweep
1.
Configure source functions:
a.
If desired, set the bias current (output current prior to the start of the
sweep) by pressing the DC key and then setting the current to the
desired value.
b.
Select the compliance display field, then set the voltage compliance as
appropriate for expected sweep parameters.
2.
Configure the sweep as follows:
a.
Press CONFIG then SWP to enter the sweep configuration menu.
b.
Select TYPE, then press ENTER.
c.
Select STAIR, and then press ENTER to choose a linear staircase
sweep.
d.
At the prompts, enter the desired START, STOP, STEP, and DELAY
values.
e.
From the CONFIGURE SWEEPS menu, select SWEEP-COUNT, press
ENTER, and then choose FINITE or INFINITE as desired.
f.
Again from the CONFIGURE SWEEPS menu, choose
SOURCE-RANGING, press ENTER, then select BEST, AUTO, or
FIXED as appropriate.
g.
From the CONFIGURE SWEEPS menu, select COMPLIANCE-ABORT,
press ENTER, then choose to abort (YES) or not to abort (NO) the
sweep if compliance is reached while the sweep is in progress.
h.
Press EXIT to return to normal display.
3.
Run sweep:
a.
Press the SWP key to arm the sweep. The output will turn on.
b.
Press TRIG to start the sweep.
c.
Press EXIT to abort the sweep before it is finished.
d.
Turn the output off by pressing the ON/OFF OUTPUT key when the
sweep is finished.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com
Содержание 6220
Страница 16: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 24: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 403: ...A Specifications Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 453: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...