Model 6220/6221 User’s Manual
SCPI Tables (Abridged)
B-3
Return to
Table B-3
Format command summary
Command
Description
Default
FORM
Reading format commands:
[:DATA] <type>
Specify data format: ASC, SRE, or DRE
ASC
:ELEM <item list>
Specify data elements: READ, TST, UNIT, RNUM, SOUR,
COMP, AVOL (or ALL or DEF)
READ,
TST
:BORD <name>
Specify byte order: NORM or SWAP.
(Note)
:SREG <name>
Select format for reading status registers: ASC, HEX, OCT or BIN.
ASC
Note: *RST default is NORM. SYST:PRES default is SWAP.
Table B-4
Output command summary
Command
Description
Default
OUTPut
Source output control:
[:STAT]?
Turn output on or off (standby).
OFF
:LTE <b>
Connect output low to earth ground (ON) or float output low (OFF).
OFF
:ISH <name>
Connect triax inner shield to OLOW (output low) or cable GUAR.
OLOW
:RESP <name>
Set the output response for 6221: FAST or SLOW.
FAST
:INT
Interlock:
:TRIP?
Returns a “0” if interlock is tripped (open) or a “1” if interlock is closed
(output enabled).
Table B-5
Sense command summary
Command
Description
Default
SENS
Commands for readings from 2182/2182A:
:DATA?
Return the latest pre-math delta reading.
:FRES?
Same as :DATA? except the reading can only be returned once.
:AVER
Averaging filter:
:TCON <name>
Select filter control: MOV or REP
MOV
:WIND <NRf>
Set filter window as % of range: 0 to 10
0.00
:COUN <NRf>
Specify filter count (size): 2 to 300
10
[:STAT] <b>
Enable or disable averaging filter.
OFF
Содержание 6220 DC
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 14: ...7 Wave Functions 6221 Only Figure 7 1 Offset example 7 4 Figure 7 2 Duty cycle 7 5 Figure 7 3 Phase marker 7 6...
Страница 16: ......
Страница 36: ...1 20 Getting Started Model 6220 6221 User s Manual Return to Section 1 topics...
Страница 131: ...6 10 Averaging Filter Math and Buffer Model 6220 6221 User s Manual Return to Section 6 topics...
Страница 148: ...A Specifications...
Страница 167: ......
Страница 169: ......
Страница 170: ......