Section 6: Clarius
Model 4200A-SCS Parameter Analyzer Reference Manual
6-146
4200A-901-01 Rev. C / February 2017
If you do not specify an instrument to be the master, the first instrument that was assigned to the step
or sweep operation mode is assigned to be the master. You can change this designation in the Test
Settings pane.
When a master is set, Clarius sets the points and step size values for the subordinate terminal to be
the same as the settings for the master terminal. You cannot change the subordinate points value for
the subordinate terminal. For list sweeps, the number of points in the list items for the subordinate is
changed to match the number of points in the master. If points are added to the master, the last point
of the subordinate list is repeated. If points are removed from the master, the same number of points
are removed from the end of the subordinate list.
You can have a dual sweep on a subordinate terminal even if the master is not set to dual sweep. In
this case, the dual sweep of the subordinate terminal has a total number of steps equal to the number
of steps in the master terminal. For example, if the master terminal is set to measure ten points, the
subordinate dual sweep will measure five points on the first side of the sweep and five points on the
second side of the dual sweep. If the master is set to an odd number of points, the subordinate
terminal will repeat the measurement of the last sweep point.
The subordinate SMUs are not automatically set for dual sweep when Dual Sweep is enabled for the
master SMU. Dual Sweep must be individually enabled for each SMU.
To specify the master sweep:
1. Select the test.
2. In the right pane, select
Test Settings
.
3. For
Sweep Master
, select the instrument that you want to designate as the master.
Configure actions
Settings for actions depend on the type of action that is selected. Actions can generate dialog boxes
to prompt test operator action, control prober movements, and manage switching. You can also
create your own actions from user libraries.
You can place actions anywhere in the project tree.
Information for actions is available in the Help pane. To open the Help pane, select the action in the
project tree and select
Configure