Model 4200A-SCS Parameter Analyzer Reference Manual
Appendix H: Using a Micromanipulator 8860 Prober
4200A-901-01 Rev. C / February 2017
H-17
8. Click
Set Reference Die
in the Setup Options window to open the Set Reference dialog box.
Figure 705: Set Reference Die button
9. Move the pins over the pads of the die to be set as the reference die.
10. Zero out the X/Y, Column, and Row (click Zero X,Y button and Zero C,R).
If you want the columns and rows to be something other than 0,0 (1,1 for instance), edit values in Set
Reference dialog box as needed before clicking
Done
.
11. Click
Done
.
Figure 706: Set Reference dialog box
Probesites Clarius project example
The following is a step-by-step procedure to configure the 8860 so the
probesites
Clarius project
executes successfully.
On the pcBridge computer:
1. Use the pcWafer program to probe a single subsite on multiple dies.
2. Start pcWafer by clicking the
pcWfr
button in the pcLaunch window. The pcWfr window appears.
Figure 707: pcWfr button