Appendix A: Using switch matrices
Model 4200A-SCS Parameter Analyzer Reference Manual
A-30
4200A-901-01 Rev. C / February 2017
Figure 544: KI 7174 Matrix Card Properties tab
Step 8. Save configuration
To save the KCon configuration:
1. Select
Save
.
Step 9. Close KCon and open Clarius
To close KCon and open Clarius:
1. To close KCon, select the close button in the upper right.
2. On the Windows desktop, select the Clarius icon.
Switch matrix control example
This example demonstrates how the
connectpins
action controls a switch matrix. You modify the
connectpins
action to connect SMU2 to a DUT, as shown in the
(on page A-
16) figure. It assumes that the switch matrix is set for row-column connections with local sense
selected. It also assumes that the matrix card properties are set as shown in
The
connectpins
action is based on the
ConnectPins
user module. Detail on
ConnectPins
is
provided in