4200-900-01 Rev. K / February 2017
Return to
1-35
Model 4200-SCS User’s Manual
Section 1: Getting Started
SENSE terminal
The SENSE terminal is a standard triaxial connector used to apply the ground unit SENSE signal
to the DUT in a remote sense application.
• The center pin is SENSE
• The inner shield is GUARD
• The outer shield is circuit COMMON
When the ground unit SENSE signal is connected to a DUT, all SMU/preamp measurements will
be made relative to this DUT connection.
COMMON terminal
The COMMON terminal is a binding post that provides access to circuit COMMON.
NOTE
Normally, a link is connected between ground unit COMMON and chassis ground,
but it may be necessary to remove the link to avoid measurement problems caused
by ground loops or electrical interference. Refer to the
for details.
Chassis ground
This binding post provides a convenient connecting point to system chassis ground for the
purpose of shielding a test fixture.
Figure 1-30
Chassis ground
INSTRUMENT
CONNECTIONS
SMU ONLY
SMU AND GNDU
GNDU
SLOT
8
S
E
N
S
E
F
O
R
C
E
C
O
M
M
O
N
SENSE LO
GUARD
SENSE LO
COMMON
COMMON
FORCE
SENSE
GUARD
Chassis (Earth) Ground
Содержание 4200-SCS
Страница 2: ......
Страница 4: ......
Страница 334: ...Index 6 4200 900 01 Rev K February 2017 Model 4200 SCS User s Manual Index...