4200-900-01 Rev. K / February 2017
Return to
1-33
Model 4200-SCS User’s Manual
Section 1: Getting Started
Figure 1-27
Ground unit
Ground unit DUT connections
shows the connections necessary to use the GNDU in conjunction with a SMU to
make full-Kelvin remote sense measurements. Similarly,
includes the preamp.
As shown in these figures, the GNDU FORCE signal provides the return path for SMU or preamp
FORCE current. For detailed information about the ground unit, SMU, and preamp connections,
refer to the
Reference Manual, Basic source-measure connections, page 4-3
.
Figure 1-28
Full-Kelvin SMU/ground unit connections
S
E
NS
E
FO
RC
E
Ground
Unit
50
k
W
C
OMMO
N
CHASSIS
32
V
max
Remo
v
able
Ground
Lin
k
Internal
4
200-SCS
connections
to
each
S
M
U
S
E
NS
E
L
O
si
g
nal
4200-SMU
or
4210-SMU
GNDU
SENSE LO
FORCE
SENSE
SENSE
FORCE
DUT