4200-900-01 Rev. K / February 2017
Return to
1-15
Model 4200-SCS User’s Manual
Section 1: Getting Started
CAUTION
The maximum allowed voltages between the preamp signals are:
•
COMMON to chassis ground: 32 V peak
•
GUARD to COMMON: 250 V peak
•
SENSE or FORCE to GUARD: 40 V peak
FORCE terminal
The FORCE terminal is a standard triaxial connector used to apply the preamp FORCE signal to
the DUT.
• The center pin is FORCE
• The inner shield is GUARD
• The outer shield is circuit COMMON
Figure 1-11
Model 4200-PA connectors
4200-P
A-1 REMO
TE P
SENSE
FORCE
P
A
RE
MP
CONT
RO
L
PreAmp
Control
Connector
Mounting
Foot
COMMON
FORCE
SENSE
GUARD
250V
PEAK
42V
PEAK
250V
PEAK
250V
PEAK
40V
PEAK
40V
PEAK
WARNING:
NO INTERNAL OPERATOR SERVICEABLE
PARTS SERVICE BY QUALIFIED
PERSONNEL ONLY.