4200-900-01 Rev. K / February 2017
Return to
3-23
Model 4200-SCS User’s Manual
Section 3: Common Device Characterization Tests
Figure 3-18
CVU Voltage Sweep output
CVU Voltage List Sweep
shows an example of a FFMO window with
CVU Voltage List Sweep
selected as the forcing function to measure Cp-Gp. The
Sweeping
test mode
must be selected for this test (see
Figure 3-19
Forcing Function: CVU Voltage List Sweep
When this test is run (see
), the following force-measure sequence
occurs:
1. The DC source goes to the PreSoak voltage of 2 V for the hold time period.
2. The DC Bias goes to the first sweep point voltage (1 V).
Содержание 4200-SCS
Страница 2: ......
Страница 4: ......
Страница 334: ...Index 6 4200 900 01 Rev K February 2017 Model 4200 SCS User s Manual Index...