3-20
Return to
4200-900-01 Rev. K / February 2017
Section 3: Common Device Characterization Tests
Model 4200-SCS User’s Manual
Status
With Status enabled (as shown in
), the following errors will be
reported in the sheet and graph tabs when a measurement fault occurs:
• ABB fails to lock
• AC voltage high overflow
• AC voltage low overflow
• AC current overflow
For more information refer to the
Reference manual, CVU measurement status,
Compensation
After making connections for the test, connection compensation must be
performed and enabled before running test. See the
Connection compensation, page 15-18
to perform connection compensation.
The CVU compensation window (shown in
) is used to enable (check)
Open
,
Short,
and
Load
compensation. The window is opened by clicking the
Compensation
button in the force-measure window.
Cable length setting is to be made from the compensation window:
•
0 M
= Disables compensation
•
1.5M
= 1.5 meters
•
3 M
= 3.0 meters
•
Custom
= Cable length coefficients measured by the user using the
Tools
>
Connection Compensation
dialog box.
Figure 3-15
CVU compensation window