Series 3700A System Switch/Multimeter Reference Manual
Section 11: TSP command reference
3700AS-901-01 Rev. D/June 2018
11-445
tspnet.write()
This function writes a string to the remote instrument.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
No
Usage
tspnet.write(
connectionID
, "
inputString
")
connectionID
The connection ID returned from
tspnet.connect()
inputString
The string to be written
Details
The
tspnet.write()
function sends
inputString
to the remote instrument. It does not wait for
command completion on the remote instrument.
The Series 3700A sends
inputString
to the remote instrument exactly as indicated. The
inputString
must contain any necessary new lines, termination, or other syntax elements needed
to complete properly.
Because
tspnet.write()
does not process output from the remote instrument, do not send
commands that generate too much output without processing the output. This command can stop
executing if there is too much unprocessed output from previous commands.
Example
tspnet.write(myID, "runscript()\r\n")
Commands the remote instrument to execute
a command or script named
runscript()
on a remote device identified in the system
as
myID
.
Also see
(on page 11-434)
(on page 11-438)