Section 11: TSP command reference
Series 3700A System Switch/Multimeter Reference Manual
11-442
3700AS-901-01 Rev. D/June 2018
tspnet.tsp.abort()
This function causes the TSP-enabled instrument to stop executing any of the commands that were sent to it.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
No
Usage
tspnet.tsp.abort(
connectionID
)
connectionID
Integer value used as a handle for other
tspnet
commands
Details
This function is appropriate only for TSP-enabled instruments.
Sends an abort command to the remote instrument.
Example
tspnet.tsp.abort(testConnection)
Stops remote instrument execution on
testConnection
.
Also see
None
tspnet.tsp.abortonconnect
This attribute contains the setting for abort on connect to a TSP-enabled instrument.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
No
Instrument reset
Recall setup
Create configuration script
Save setup
1 (enable)
Usage
tspnet.tsp.abortonconnect =
value
value
= tspnet.tsp.abortonconnect
value
▪
Enable:
1
▪
Disable:
0
Details
This setting determines if the instrument sends an abort message when it attempts to connect to a
TSP-enabled instrument using the
tspnet.connect()
function.
When you send the abort command on an interface, it causes any other active interface on that
instrument to close. If you do not send an abort command (or if
tspnet.tsp.abortonconnect
is
set to 0) and another interface is active, connecting to a TSP-enabled remote instrument results in a
connection. However, the instrument will not respond to subsequent reads or executes because
control of the instrument is not obtained until an abort command has been sent.