Series 3700A System Switch/Multimeter Reference Manual
Section 11: TSP command reference
3700AS-901-01 Rev. D/June 2018
11-289
localnode.prompts4882
This attribute enables and disables the generation of prompts for IEEE Std 488.2 common commands.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Power cycle
Not saved
1 (enabled)
Usage
prompting
= localnode.prompts4882
localnode.prompts4882 =
prompting
prompting
IEEE Std 488.2 prompting mode:
▪
Disable prompting:
0
▪
Enable prompting:
1
Details
When this attribute is enabled, the IEEE Std 488.2 common commands generate prompts if
prompting is enabled with the
localnode.prompts
attribute. If
localnode.prompts4882
is
enabled, limit the number of
*trg
commands sent to a running script to 50 regardless of the setting
of the
localnode.prompts
attribute.
When this attribute is disabled, IEEE Std 488.2 common commands will not generate prompts. When
using the
*trg
command with a script that executes
trigger.wait()
repeatedly, disable
prompting to avoid problems associated with the command interface input queue filling.
Example
localnode.prompts4882 = 0
Disables IEEE Std 488.2 common command prompting.
Also see
localnode.reset()
This function resets the local node instrument.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
localnode.reset()
Details
If you want to reset a specific instrument or a subordinate node, use the
node[
X
].reset()
command.