Model 2790 SourceMeter® Switch System Reference Manual
Miscellaneous calibration commands
Miscellaneous commands perform various calibration functions such as programming the
calibration code and date. These commands are discussed in detail in the following
paragraphs.
:CODE
(:CALibration:PROTected:CODE)
Purpose
To program the calibration code or password so that you can perform the
calibration procedures.
Format
:cal:prot:code '<char_string>'
Parameters
Up to a 8-character string including letters and numbers.
Description
The
:CODE
command enables the Model 2790 calibration procedures
when performing these procedures over the bus. In general, this com-
mand must be sent to the unit before sending any other comprehensive
or manufacturing calibration command. The default calibration code is
KI002790.
klqb
The
:CODE
command should be sent only once before performing either the
comprehensive or factory calibration. Do not send
:CODE
before each calibra-
tion step.
To change the code, first send the present code, then send the new code.
The code parameter must be enclosed in single quotes.
Example
:CAL:PROT:CODE 'KI002790'
Send default code of KI002790.
:COUNt?
(:CALibration:PROTected:COUNt?)
Purpose
To determine how many times the Model 2790 has been calibrated.
Format
:cal:prot:coun?
Response
<n> Calibration count.
Description
The
:COUNt?
command allows you to determine how many times the
Model 2790 has been calibrated. Use the :COUNt? command to help
you monitor for unauthorized calibration procedures.
Example
:CAL:PROT:COUN?
Request calibration count.
Содержание 2790
Страница 3: ......
Страница 7: ......
Страница 105: ...2 48 Model 2790 SourceMeter Switch System Reference Manual...
Страница 245: ...7 40 Model 2790 SourceMeter Switch System Reference Manual...
Страница 265: ...8 20 Model 2790 SourceMeter Switch System Reference Manual...
Страница 305: ...10 22 Model 2790 SourceMeter Switch System Reference Manual...
Страница 328: ...12 CommonCommands...
Страница 336: ...13 SCPISignalOriented MeasurementCommands...
Страница 345: ...13 10 Model 2790 SourceMeter Switch System Reference Manual...
Страница 355: ...14 10 Model 2790 SourceMeter Switch System Reference Manual...
Страница 356: ...15 SCPIReferenceTables...
Страница 419: ...16 38 Model 2790 SourceMeter Switch System Reference Manual...
Страница 461: ...A 2 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Страница 462: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 3...
Страница 463: ...A 4 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Страница 464: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 5...
Страница 465: ...A 6 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Страница 466: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 7...
Страница 471: ...A 12 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Страница 472: ...B StatusandErrorMessages...
Страница 479: ...B 8 Model 2790 SourceMeter Switch System Reference Manual...
Страница 493: ...C 14 Model 2790 SourceMeter Switch System Reference Manaul...
Страница 509: ...E 6 Model 2790 SourceMeter Switch System Reference Manual...
Страница 510: ...F IEEE 488BusOverview...
Страница 525: ...F 16 Model 2790 SourceMeter Switch System Reference Manual...
Страница 526: ...G CalibrationReference...