3-16
Model 2790 SourceMeter
®
Swich System Reference Manual
AMPS fuse replacement (front panel AMPS input)
t^okfkd
Make sure the instrument is disconnected from the power line and
other equipment before replacing the AMPS fuse.
1.
Turn off the power and disconnect the power line and test leads.
2.
From the front panel, gently push in the AMPS fuse holder with a flat-blade
screwdriver and rotate the fuse holder one-quarter turn counterclockwise.
3.
Remove the fuse and replace it with the same type (3A, 250V, fast-blow,
5
×
20mm). The Keithley part number is FU-99-1.
`^rqflk
Do not use a fuse with a higher current rating than specified or
instrument damage may occur. If the instrument repeatedly blows
fuses, locate and correct the cause of the trouble before replacing the
fuse.
4.
Install the new fuse by reversing the above procedure.
klqb
For the Model 7702 switching module, there are solder mount amps fuses. See
of this manual for fuse replacement information.
Resistance measurements (
Ω
2 and
Ω
4)
The Model 2790 uses the constant-current method to measure resistance from 100
Ω
to
1M
Ω
. The Model 2790 sources a constant current (I) to the resistance and measures the
voltage (V). Resistance (R) is then calculated (and displayed) using the known current and
measured voltage (R = V/I). For the 10M
Ω
and 100M
Ω
ranges, the ratiometric method is
used to measure resistance.
Standard resistance measurements
— The Model 2790 can make resistance measure-
ments from 100µ
Ω
to 120M
Ω
. For resistances >1k
Ω
, the 2-wire (
Ω
2) method is typically
used for measurements. For resistances
≤1
k
Ω
, the 4-wire (
Ω
4) measurement method
should be used to cancel the effect of test lead (and channel path) resistances.
Offset-compensated ohms (OCOMP)
— The presence of thermal EMFs (voltages) can
adversely affect low-resistance measurement accuracy. To overcome these unwanted off-
set voltages, you can use offset compensated ohms on the 100
Ω,
1k
Ω,
and 10k
Ω
ranges
for the
Ω
4 function.
Содержание 2790
Страница 3: ......
Страница 7: ......
Страница 105: ...2 48 Model 2790 SourceMeter Switch System Reference Manual...
Страница 245: ...7 40 Model 2790 SourceMeter Switch System Reference Manual...
Страница 265: ...8 20 Model 2790 SourceMeter Switch System Reference Manual...
Страница 305: ...10 22 Model 2790 SourceMeter Switch System Reference Manual...
Страница 328: ...12 CommonCommands...
Страница 336: ...13 SCPISignalOriented MeasurementCommands...
Страница 345: ...13 10 Model 2790 SourceMeter Switch System Reference Manual...
Страница 355: ...14 10 Model 2790 SourceMeter Switch System Reference Manual...
Страница 356: ...15 SCPIReferenceTables...
Страница 419: ...16 38 Model 2790 SourceMeter Switch System Reference Manual...
Страница 461: ...A 2 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Страница 462: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 3...
Страница 463: ...A 4 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Страница 464: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 5...
Страница 465: ...A 6 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Страница 466: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 7...
Страница 471: ...A 12 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Страница 472: ...B StatusandErrorMessages...
Страница 479: ...B 8 Model 2790 SourceMeter Switch System Reference Manual...
Страница 493: ...C 14 Model 2790 SourceMeter Switch System Reference Manaul...
Страница 509: ...E 6 Model 2790 SourceMeter Switch System Reference Manual...
Страница 510: ...F IEEE 488BusOverview...
Страница 525: ...F 16 Model 2790 SourceMeter Switch System Reference Manual...
Страница 526: ...G CalibrationReference...