1-20
Getting Started
Model 2750 Multimeter/Switch System User’s Manual
Saving and restoring setups
Saving a user setup
1.
Configure Model 2750 for the desired measurement application.
2.
Press SHIFT and then SAVE to access the save setup menu.
3.
Press
to place the cursor on the present setup (SAV0, SAV1, or SAV2).
4.
Use the
Δ
or
∇
key to display the desired setup and press ENTER. The instrument
returns to the normal measurement state.
WARNING
If you make firmware upgrades you will lose all your saved settings.
Saving a power-on setup
1.
Configure Model 2750 for the desired measurement application.
2.
Press SHIFT and then SAVE to access the save setup menu.
3.
Press the
Δ
key to display the present power-on (PWR-ON) setup: FACT, *RST,
SAV0, SAV1, or SAV2.
4.
Press
to place the cursor on the present power-on setup.
5.
Use the
Δ
or
∇
key to display the desired setup and press ENTER. The instrument
returns to the normal measurement state.
Restoring a setup
1.
Press SHIFT and then SETUP to access the restore setup menu.
2.
Press
to place the cursor on the present RESTORE setup (FACT, *RST, SAV0,
SAV1, or SAV2).
3.
Use the
Δ
or
∇
key to display the desired setup and press ENTER. The instrument
returns to the normal measurement state.
NOTE
If the settings for a user setup or power-on setup do not match the switching
module types presently installed in the Model 2750, error +520 (Saved setup
scancard mismatch) occurs when the setup is recalled. The scan list will reset to
the factory defaults and all channels will open. However, the saved setup is still
retained in memory and can be restored when the matching switching module is
later installed.
2750-900-01.book Page 20 Wednesday, August 3, 2011 7:56 AM
Содержание 2750
Страница 3: ...2750 900 01 book Page 2 Wednesday August 3 2011 7 56 AM...
Страница 7: ...2750 900 01 book Page 4 Wednesday August 3 2011 7 56 AM...
Страница 17: ...2750 900 01 book Page 10 Wednesday August 3 2011 7 56 AM...
Страница 204: ...Model 2750 Multimeter Switch System User s Manual Buffer 6 15 2750 900 01 book Page 15 Wednesday August 3 2011 7 56 AM...
Страница 205: ...6 16 Buffer Model 2750 Multimeter Switch System User s Manual 2750 900 01 book Page 16 Wednesday August 3 2011 7 56 AM...
Страница 324: ...12 CommonCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 332: ...13 SCPISignalOriented MeasurementCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 348: ...15 SCPIReferenceTables 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 374: ...B Model7700ConnectionGuide 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 386: ...C StatusandErrorMessages 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 394: ...D SignalProcessing SequenceandDataFlow 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 408: ...E MeasurementConsiderations 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 428: ...G IEEE 488BusOverview 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...