Model 2750 Multimeter/Switch System User’s Manual
SCPI Signal Oriented Commands
13-7
MEASure:<function>? [<rang>], [<res>], [<clist>]
MEASure:VOLTage[:DC]? [<rang>], [<res>], [<clist>]
Measure DCV
MEASure:VOLTage:AC? [<rang>], [<res>], [<clist>]
Measure ACV
MEASure:CURRent[:DC]? [<rang>], [<res>], [<clist>]
Measure DCI
MEASure:CURRent:AC? [<rang>], [<res>], [<clist>]
Measure ACI
MEASure:RESistance? [<rang>], [<res>], [<clist>]
Measure
Ω
2
MEASure:FRESistance? [<rang>], [<res>], [<clist>]
Measure
Ω
4
MEASure:FREQuency? [<rang>], [<res>], [<clist>]
Measure FREQ
MEASure:PERiod? [<rang>], [<res>], [<clist>]
Measure PERIOD
MEASure:TEMPerature? [<clist>]
Measure TEMP
MEASure:CONTinuity? [<clist>]
Measure CONT
Parameters
<rang>
= Range parameter for the specified function. For example, for
DCV, range parameter value 10 selects the 10V range. See the
“NOTES” that follow
for additional information.
<res> = 0.1
i.e., 100.0 V (3
H
digits)
0.01
i.e., 10.00 V (3
H
digits)
0.001
i.e., 1.000 V (3
H
digits)
0.0001
i.e., 1.0000 V (4
H
digits)
0.00001
i.e., 1.00000 V (5
H
digits)
0.000001
i.e., 1.000000 V (6
H
digits)
The resolution of the <res> parameter value and the selected
range sets the number of display digits. As shown above, with
the 100V range selected and <res> = 0.1, a 100V reading will
be displayed as 100.0 V (3
H
digits).
The display will default to 3
H
digits when using parameter
values that attempt to set the display below 3
H
digits. For
example, a 10V reading using <res> = 0.1 for the 10V range is
displayed as 10.00 V, not 10.0 V.
A command using parameter values that attempt to set the
display above 7
H
digits is ignored, and generates error -221
(settings conflict error).
for additional
information.
<clist>
=
Single channel only. When included, this is the channel to be
closed and measured.
Description
The MEASure? command combines all of the other signal oriented
measurement commands to perform a “one-shot” measurement and
acquire the reading. If the <clist> parameter is included, the specified
channel will close before performing the measurement.
2750-900-01.book Page 7 Wednesday, August 3, 2011 7:56 AM
Содержание 2750
Страница 3: ...2750 900 01 book Page 2 Wednesday August 3 2011 7 56 AM...
Страница 7: ...2750 900 01 book Page 4 Wednesday August 3 2011 7 56 AM...
Страница 17: ...2750 900 01 book Page 10 Wednesday August 3 2011 7 56 AM...
Страница 204: ...Model 2750 Multimeter Switch System User s Manual Buffer 6 15 2750 900 01 book Page 15 Wednesday August 3 2011 7 56 AM...
Страница 205: ...6 16 Buffer Model 2750 Multimeter Switch System User s Manual 2750 900 01 book Page 16 Wednesday August 3 2011 7 56 AM...
Страница 324: ...12 CommonCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 332: ...13 SCPISignalOriented MeasurementCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 348: ...15 SCPIReferenceTables 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 374: ...B Model7700ConnectionGuide 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 386: ...C StatusandErrorMessages 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 394: ...D SignalProcessing SequenceandDataFlow 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 408: ...E MeasurementConsiderations 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 428: ...G IEEE 488BusOverview 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...