7-26
Scanning
Model 2750 Multimeter/Switch System User’s Manual
Remote programming — scanning
NOTE
Scanning examples (remote programming and front panel operation) are pro-
vided at the end of this section.
Trigger model
The trigger model for bus operation is shown in
. Bus operation is similar to
front panel SCAN operation, with the following significant differences:
Idle
— The instrument goes into the idle state (measurements halted) after the last scan
channel is measured. For front panel operation, the instrument stays in idle until the next
scan is started. For bus operation, the instrument will not stay in idle unless continuous ini-
tiation is disabled. There are two commands to disable continuous initiation:
INITitate:CONTinuous OFF
' Disable continuous initiation.
*RST
' Restore *RST defaults.
The instrument will remain in idle until it receives an initiate command. Typical com-
mands to initiate one scan cycle include:
INITiate
' Initiate one scan cycle.
READ?
' Initiate one scan cycle and request “sample” readings.
More information on using these commands is provided by Reference c that follows
Control sources
— For bus operation, there are two additional control sources: Bus and
Manual. For the Bus control source, scan operation is controlled by bus triggers (i.e.,
*TRG) or by using the TRIG key. For the Manual control source, event detection is con-
trolled solely by the TRIG key. Note that the instrument has to be in local in order to use
the TRIG key. The LOCAL key takes the instrument out of remote.
Trigger and sample counters
— For front panel SCAN operation, the number of chan-
nels in the scan list and the programmed reading count automatically sets the trigger and
sample counters. For remote operation, these two counters are set by the
TRIGGer:COUNt and SAMPle:COUNt commands.
NOTE
To set sample count >1, continuous initiation must be disabled (see “
”).
Note that only sample count readings are stored in the buffer. See
for
detailed information on the trigger model.
2750-900-01.book Page 26 Wednesday, August 3, 2011 7:56 AM
Содержание 2750
Страница 3: ...2750 900 01 book Page 2 Wednesday August 3 2011 7 56 AM...
Страница 7: ...2750 900 01 book Page 4 Wednesday August 3 2011 7 56 AM...
Страница 17: ...2750 900 01 book Page 10 Wednesday August 3 2011 7 56 AM...
Страница 204: ...Model 2750 Multimeter Switch System User s Manual Buffer 6 15 2750 900 01 book Page 15 Wednesday August 3 2011 7 56 AM...
Страница 205: ...6 16 Buffer Model 2750 Multimeter Switch System User s Manual 2750 900 01 book Page 16 Wednesday August 3 2011 7 56 AM...
Страница 324: ...12 CommonCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 332: ...13 SCPISignalOriented MeasurementCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 348: ...15 SCPIReferenceTables 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 374: ...B Model7700ConnectionGuide 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 386: ...C StatusandErrorMessages 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 394: ...D SignalProcessing SequenceandDataFlow 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 408: ...E MeasurementConsiderations 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 428: ...G IEEE 488BusOverview 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...