Model 2750 Multimeter/Switch System User’s Manual
Range, Digits, Rate, Bandwidth, and Filter
4-21
Commands
1, 4
Description
5
Default
Ω
4 filter commands
[SENSe[1]]
Optional root command.
:FRESistance:AVERage:TCONtrol <name>
Select filter type; <name> = MOVing or
REPeat.
(Note 2)
:FRESistance:AVERage:WINDow <NRf>
Set filter window in %; <NRf> = 0 to 10.
0.1
:FRESistance:AVERage:COUNt <n> [, clist]
Specify filter count; <n> = 1 to 100.
10
:FRESistance:AVERage:STATe <b> [, clist]
Enable or disable the filter.
(Note 3)
TEMP filter commands
[SENSe[1]]
Optional root command.
:TEMPerature:AVERage:TCONtrol <name>
Select filter type; <name> = MOVing or
REPeat.
(Note 2)
:TEMPerature:AVERage:WINDow <NRf>
Set filter window in %; <NRf> = 0 to 10.
0.1
:TEMPerature:AVERage:COUNt <n> [, clist]
Specify filter count; <n> = 1 to 100.
10
:TEMPerature:AVERage:STATe <b> [, clist]
Enable or disable the filter.
(Note 3)
Channel list parameter:
<clist> = (@SCH)
where: S =
Mainframe
slot
number (1, 2, 3, 4 or 5)
CH = Switching module channel number (must be 2 digits)
Examples:
(@101) = Slot 1, Channel 1
(@101, 203) = Slot 1, Channel 1 and Slot 2, Channel 3
(@101:110) = Slot 1, Channels 1 through 10
Notes:
1. The <clist> parameter is used to configure one or more channels for a scan. Each channel in the <clist> must be set to the function
specified by the filter command. If not, a conflict error (-221) will occur. For example, VOLTage:AVERage:STATe ON, (@101) is
only valid if scan channel 101 is set for the DCV function.
2. REPeat is the *RST default, and MOVing is the SYSTem:PRESet default. From the front panel, the factory default is MOVing.
3. OFF is the *RST default, and ON is the SYSTem:PRESet default.
4. [:DC] is optional for the commands to set DCV and DCI filter.
5. Filter window — Parameter value 0 for the WINDow commands sets the filter window to NONE.
Table 4-6 (continued)
Filter commands
2750-900-01.book Page 21 Wednesday, August 3, 2011 7:56 AM
Содержание 2750
Страница 3: ...2750 900 01 book Page 2 Wednesday August 3 2011 7 56 AM...
Страница 7: ...2750 900 01 book Page 4 Wednesday August 3 2011 7 56 AM...
Страница 17: ...2750 900 01 book Page 10 Wednesday August 3 2011 7 56 AM...
Страница 204: ...Model 2750 Multimeter Switch System User s Manual Buffer 6 15 2750 900 01 book Page 15 Wednesday August 3 2011 7 56 AM...
Страница 205: ...6 16 Buffer Model 2750 Multimeter Switch System User s Manual 2750 900 01 book Page 16 Wednesday August 3 2011 7 56 AM...
Страница 324: ...12 CommonCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 332: ...13 SCPISignalOriented MeasurementCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 348: ...15 SCPIReferenceTables 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 374: ...B Model7700ConnectionGuide 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 386: ...C StatusandErrorMessages 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 394: ...D SignalProcessing SequenceandDataFlow 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 408: ...E MeasurementConsiderations 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 428: ...G IEEE 488BusOverview 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...