Model 2750 Multimeter/Switch System User’s Manual
Basic DMM Operation
3-21
The offset-compensated ohms reading is then calculated as follows:
Offset-compensated ohms reading =
Δ
V/
Δ
I
where:
Δ
V = V2 - V1
Δ
I = I2 - I1
V1 is the voltage measurement with the current source at its normal level.
V2 is the voltage measurement using the lowest current source setting.
The above 2-point measurement process and reading calculation eliminates the resistance
contributed by the presence of V
EMF
.
Enabling/disabling offset-compensated ohms
Offset-compensated ohms is enabled by pressing SHIFT and then OCOMP. When
enabled, the OCOMP annunciator is on. Offset-compensated ohms is disabled by again
pressing SHIFT and then OCOMP.
Performing offset-compensated ohms measurements
Offset-compensated ohms can only be performed on the
Ω
4 function using the 1
Ω
, 10
Ω
,
100
Ω
, 1k
Ω
, or 10k
Ω
range. Make sure you use 4-wire connections to the DUT (see
NOTE
Make sure the INPUTS switch is in the correct position. To use front panel
inputs, it must be in the “F” (out) position. For switching modules, it must be in
the “R” (in) position.
1.
If a switching channel is presently closed (displayed), press OPEN to open it.
2.
Select the 4-wire ohms measurement function by pressing
Ω
4, and enable offset
compensated ohms by pressing SHIFT and then OCOMP (OCOMP annunciator
turns on).
3.
Use the RANGE up and down keys to select the 1
Ω
, 10
Ω
, 100
Ω
, 1k
Ω
, or 10k
Ω
range, or press AUTO to enable auto range. If using auto range, offset-compen-
sated ohms measurements will not be performed if the instrument goes to the
100k
Ω
or higher range.
4.
Perform steps 4 through 8 of the “
Standard resistance measurements
” procedure.
2750-900-01.book Page 21 Wednesday, August 3, 2011 7:56 AM
Содержание 2750
Страница 3: ...2750 900 01 book Page 2 Wednesday August 3 2011 7 56 AM...
Страница 7: ...2750 900 01 book Page 4 Wednesday August 3 2011 7 56 AM...
Страница 17: ...2750 900 01 book Page 10 Wednesday August 3 2011 7 56 AM...
Страница 204: ...Model 2750 Multimeter Switch System User s Manual Buffer 6 15 2750 900 01 book Page 15 Wednesday August 3 2011 7 56 AM...
Страница 205: ...6 16 Buffer Model 2750 Multimeter Switch System User s Manual 2750 900 01 book Page 16 Wednesday August 3 2011 7 56 AM...
Страница 324: ...12 CommonCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 332: ...13 SCPISignalOriented MeasurementCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 348: ...15 SCPIReferenceTables 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 374: ...B Model7700ConnectionGuide 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 386: ...C StatusandErrorMessages 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 394: ...D SignalProcessing SequenceandDataFlow 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 408: ...E MeasurementConsiderations 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 428: ...G IEEE 488BusOverview 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...