Model 2750 Multimeter/Switch System Service Manual
Accuracy and Optimization
A-5
Sp
ecif
ica
tions
Optimizing measurement speed
The configurations listed below assume that the multimeter has had factory setups
restored.
DC voltage, DC current, and resistance:
•
Select 3-1/2 digits, 0.01 PLC, filter OFF, fixed range.
AC voltage and AC current:
•
Select 3-1/2 digits, 0.01 PLC, filter OFF, fixed range.
Temperature:
•
Select 3-1/2 digits, 0.01 PLC, filter OFF.
For all functions, turn off the display and autozero, and set the trigger delay to zero. Use
the :SAMPle:COUNt and READ? bus commands.