Model 2701 User’s Manual
Remote Operations
10-3
Separate function setups
A few settings from the front panel are global. That is, the setting on one function also
applies to the other functions. For example, if you set DCV for 3½ digits, all the other
functions will also be set to 3½ digits. Using remote programming, each function can have
its own unique setup. For example, DCV can be set to 3½ digits, ACI can be set to
4½ digits, and DCI can be set to 5½ digits.
Global settings from the front panel that can be set separately using remote programming
include digits, rate, and filter configuration (except on/off, which can be set separately).
NOTE
Do not confuse function setups with scan channel setups. For scan channels,
separate settings for digits, rate, and filter configuration can be set from either
the front panel or remote programming. See Section 7 for details on scan
channel setup.
DCV input divider
Using remote programming, you can enable the DCV input divider for the 100mV, 1V,
and 10V ranges. When enabled, the input resistance for these DCV ranges are reduced to
10M
Ω
. See Section 3 for details on the DCV input divider.
Multiple channel operation
For normal system channel operation, when one measurement channel is closed, the
previous measurement channel opens. With the use of the ROUTe:MULTiple commands,
you gain independent control of all switching module channels, including the relays that
connect the input signal to the DMM. See Section 2 for details.
System commands
Interface
SYStem subsystem commands are used to select and configure the interface for remote
programming. System commands for the Ethernet are provided in Table 10-2, while
system commands for the RS-232 interface are provided in Table 10-3.
NOTE
The Model 2701 has an internal web page. From this web page, you can send
commands to the 2701 and take readings. See “Internal web page” for details.
2701-900-01.book Page 3 Wednesday, August 3, 2011 9:43 AM
Содержание 2701
Страница 344: ...12 CommonCommands 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 352: ...13 SCPISignalOriented MeasurementCommands 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 366: ...15 SCPIReferenceTables 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 393: ...A Specifications 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 399: ...B Model7700ConnectionGuide 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 410: ...C StatusandErrorMessages 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 417: ...D SignalProcessing SequenceandDataFlow 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 430: ...E MeasurementConsiderations 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 450: ...G KE2700InstrumentDriver Examples 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...