3-48
Basic DMM Operation
Model 2701 User’s Manual
Frequency and period measurement procedure
NOTE
Make sure the INPUTS switch is in the correct position. To use front panel
inputs, it must be in the “F” (out) position. For switching modules, it must be in
the “R” (in) position.
1.
If a switching channel is presently closed (displayed), press OPEN to open it.
2.
Perform one of the following steps to select the function:
•
Press FREQ to perform frequency measurements.
•
Press SHIFT and then FREQ to perform period measurements.
3.
Use the RANGE and keys to select a measurement range consistent with the
expected AC voltage. Details on range are provided in Section 4.
4.
Apply the AC voltage(s) to be measured.
CAUTION
Do not apply more than the maximum input levels indicated in Figure 3-17
and Figure 3-18 or instrument damage may occur.
5.
If using a switching module, perform the following steps to close the desired
channel:
a.
Press the CLOSE key.
b.
Use
,
, , and to key in the channel number and press ENTER. The
previously closed channel (if there is one) will open and the specified channel
will close.
NOTE
While in the normal measurement state, you can use the
and
keys to close
channels. In general, each key press will open the presently closed channel and
then close the next higher or lower channel.
6.
Observe the displayed reading. If the “OVERFLOW” message is displayed, select
a higher range until a normal reading is displayed. Use the lowest possible range
for the best resolution.
7.
To measure other switching channels, repeat steps 5 and 6.
8.
When finished, press OPEN if there is a channel closed.
Continuity testing
The instrument can test continuity using the 2-wire 1k
Ω
range. After selecting continuity,
you will be prompted to enter the threshold resistance level (1 to 1000
Ω
). When the
measured circuit is below the set threshold level, the instrument will beep and display the
resistance readings. When the measured circuit is above the threshold level, the instrument
2701-900-01.book Page 48 Wednesday, August 3, 2011 9:43 AM
Содержание 2701
Страница 344: ...12 CommonCommands 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 352: ...13 SCPISignalOriented MeasurementCommands 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 366: ...15 SCPIReferenceTables 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 393: ...A Specifications 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 399: ...B Model7700ConnectionGuide 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 410: ...C StatusandErrorMessages 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 417: ...D SignalProcessing SequenceandDataFlow 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 430: ...E MeasurementConsiderations 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 450: ...G KE2700InstrumentDriver Examples 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...