Model 2701 User’s Manual
Basic DMM Operation
3-23
Cable leakage
For high resistance measurements in a high humidity environment, use Teflon™ insulated
cables to minimize errors due to cable leakage.
Standard resistance measurements
NOTE
Make sure the INPUTS switch is in the correct position. To use front panel
inputs, it must be in the “F” (out) position. For switching modules, it must be in
the “R” (in) position.
Perform the following steps to measure resistance:
1.
If a switching channel is presently closed (displayed), press OPEN to open it.
2.
Select the ohms measurement function by pressing
Ω
2 or
Ω
4.
3.
Use the RANGE and keys to select a measurement range consistent with the
expected resistance or press AUTO to select autoranging (AUTO annunciator turns
on). Details on range are provided in Section 4.
4.
Connect the resistance(s) to be measured.
CAUTION
Front panel inputs — Do not apply more than 1000V peak between INPUT
HI and LO or instrument damage may occur.
Model 7700 switching module — Do not apply more than 300V DC or
300V RMS (425V peak) for AC waveforms between input high (H) or
input low (L) or switching module damage may occur.
5.
If using a switching module, perform the following steps to close the desired
channel. Keep in mind that for
Ω
4 measurements, you will close the primary
(INPUT) channel (1 through 10). The paired channel will close automatically.
a.
Press the CLOSE key.
b.
Use
,
, , and , to key in the channel number and press ENTER. The
previously closed channel(s) (if any) will open, and the specified channel (or
channel pair) will close.
NOTE
While in the normal measurement state, you can use the
and keys to close
channels. In general, each key press will open the presently closed channel and
then close the next higher or lower channel.
6.
Observe the displayed reading. If the “OVERFLOW” message is displayed, select
a higher range until a normal reading is displayed (or press AUTO for
autoranging). For manual ranging, use the lowest possible range for the best
resolution.
7.
To measure other switching channels, repeat steps 5 and 6.
8.
When finished, press OPEN if there is a channel closed.
2701-900-01.book Page 23 Wednesday, August 3, 2011 9:43 AM
Содержание 2701
Страница 344: ...12 CommonCommands 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 352: ...13 SCPISignalOriented MeasurementCommands 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 366: ...15 SCPIReferenceTables 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 393: ...A Specifications 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 399: ...B Model7700ConnectionGuide 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 410: ...C StatusandErrorMessages 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 417: ...D SignalProcessing SequenceandDataFlow 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 430: ...E MeasurementConsiderations 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 450: ...G KE2700InstrumentDriver Examples 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...