Model 2701 User’s Manual
Basic DMM Operation
3-19
6.
Observe the displayed reading. If the “OVERFLOW” message is displayed, select
a higher range until a normal reading is displayed (or press AUTO for
autoranging). For manual ranging, use the lowest possible range for the best
resolution.
7.
To measure another amps channel, repeat steps 5 and 6.
8.
When finished, press OPEN if there is a channel closed.
NOTE
When you have an amps-only channel closed, you cannot select a non-amps
function. For example, if channel 21 of the Model 7700 is closed, you cannot
select the DCV function (“INVALID FUNC” displayed).
AMPS fuse replacement (front panel AMPS input)
WARNING
Make sure the instrument is disconnected from the power line and other
equipment before replacing the AMPS fuse.
1.
Turn off the power and disconnect the power line and test leads.
2.
From the front panel, gently push in the AMPS fuse holder with a flat blade
screwdriver and rotate the fuse holder one-quarter turn counterclockwise.
3.
Remove the fuse and replace it with the same type (3A, 250V, fast-blow,
5
×
20mm). The Keithley part number is FU-99-1.
CAUTION
Do not use a fuse with a higher current rating than specified or instrument
damage may occur. If the instrument repeatedly blows fuses, locate and
correct the cause of the trouble before replacing the fuse.
4.
Install the new fuse by reversing the procedure above.
NOTE
For the Model 7700 switching module and other similar modules that support
the amps function, there are solder mount amps fuses. See the Model 2701
Service Manual for fuse replacement information.
2701-900-01.book Page 19 Wednesday, August 3, 2011 9:43 AM
Содержание 2701
Страница 344: ...12 CommonCommands 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 352: ...13 SCPISignalOriented MeasurementCommands 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 366: ...15 SCPIReferenceTables 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 393: ...A Specifications 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 399: ...B Model7700ConnectionGuide 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 410: ...C StatusandErrorMessages 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 417: ...D SignalProcessing SequenceandDataFlow 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 430: ...E MeasurementConsiderations 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...
Страница 450: ...G KE2700InstrumentDriver Examples 2701 900 01 book Page 1 Wednesday August 3 2011 9 43 AM...