Model 2700 Multimeter/Switch System User’s Manual
Getting Started
1-21
Saving and restoring setups
Saving a user setup
1.
Configure Model 2700 for the desired measurement application.
2.
Press SHIFT and then SAVE to access the save setup menu.
3.
Press
to place the cursor on the present setup (SAV0, SAV1, SAV2, SAV3).
4.
Use the
Δ
or
∇
key to display the desired setup and press ENTER. The instrument
returns to the normal measurement state.
Saving a power-on setup
1.
Configure Model 2700 for the desired measurement application.
2.
Press SHIFT and then SAVE to access the save setup menu.
3.
Press the
Δ
key to display the present power-on (PWR-ON) setup; FACT, *RST,
SAV0, SAV1, SAV2, or SAV3.
4.
Press
to place the cursor on the present power-on setup.
5.
Use the
Δ
or
∇
key to display the desired setup and press ENTER. The instrument
returns to the normal measurement state.
Restoring a setup
1.
Press SHIFT and then SETUP to access the restore setup menu.
2.
Press
to place the cursor on the present RESTORE setup (FACT, *RST, SAV0,
SAV1, SAV2, or SAV3).
3.
Use the
Δ
or
∇
key to display the desired setup and press ENTER. The instrument
returns to the normal measurement state.
NOTE
If the settings for a user setup or power-on setup do not match the switching
module types presently installed in the Model 2700, error +520 (Saved setup
scancard mismatch) occurs when the setup is recalled. The scan list will reset to
the factory defaults and all channels will open. However, the saved setup is still
retained in memory and can be restored when the matching switching module is
later installed.
Содержание 2700
Страница 3: ......
Страница 7: ......
Страница 197: ...5 24 Rel Math Ratio Channel Average dB Model 2700 Multimeter Switch System User s Manual...
Страница 213: ...6 16 Buffer Model 2700 Multimeter Switch System User s Manual...
Страница 289: ...9 18 Limits and Digital I O Model 2700 Multimeter Switch System User s Manual...
Страница 335: ...11 24 Status Structure Model 2700 Multimeter Switch System User s Manual...
Страница 336: ...12 CommonCommands...
Страница 344: ...13 SCPISignalOriented MeasurementCommands...
Страница 353: ...13 10 SCPI Signal Oriented Commands Model 2700 Multimeter Switch System User s Manual...
Страница 363: ...14 10 FORMat and Misc SYSTem Commands Model 2700 Multimeter Switch System User s Manual...
Страница 364: ...15 SCPIReferenceTables...
Страница 391: ...A 2 Specifications Model 2700 Multimeter Switch System User s Manual...
Страница 392: ...Model 2700 Multimeter Switch System User s Manual Specifications A 3...
Страница 393: ...A 4 Specifications Model 2700 Multimeter Switch System User s Manual...
Страница 394: ...Model 2700 Multimeter Switch System User s Manual Specifications A 5...
Страница 395: ...A 6 Specifications Model 2700 Multimeter Switch System User s Manual...
Страница 400: ...B Model7700ConnectionGuide...
Страница 411: ...B 12 Model 7700 Connection Guide Model 2700 Multimeter Switch System User s Manual...
Страница 412: ...C StatusandErrorMessages...
Страница 419: ...C 8 Status and Error Messages Model 2700 Multimeter Switch System User s Manual...
Страница 420: ...D SignalProcessing SequenceandDataFlow...
Страница 433: ...D 14 Signal Processing Sequence and Data Flow Model 2700 Multimeter Switch System User s Manual...
Страница 434: ...E MeasurementConsiderations...
Страница 443: ...E 10 Measurement Considerations Model 2700 Multimeter Switch System User s Manual...
Страница 454: ...G IEEE 488BusOverview...
Страница 469: ...G 16 IEEE 488 Bus Overview Model 2700 Multimeter Switch System User s Manual...
Страница 470: ...H KE2700InstrumentDriver Examples...
Страница 489: ...H 20 KE2700 Instrument Driver Examples Model 2700 Multimeter Switch System User s Manual...
Страница 499: ......