Model 2700 Multimeter/Switch System User’s Manual
Scanning
7-21
Saving setup
Up to four instrument setups can be saved in memory using the SHIFT > SAVE menu
(SAV0, SAV1, SAV2, or SAV3). A user-saved setup can also be used as the power-on
setup. A user-saved setup can be restored from the SHIFT > SETUP menu. Details on
user-setups are covered in
.
Auto scan
When auto scan is enabled, the scan operation is saved in memory. If power to the
Model 2700 is interrupted, the scan will resume when power is restored. With auto scan
enabled, the last scan setup becomes the power-on setup. It takes precedence over the
factory, *RST, or user-saved power-on setup.
Perform the following steps to set auto scan.
1.
While in the normal measurement state, press SHIFT and then SETUP.
2.
Use the
Δ
and
∇
keys to display the auto scan (AUTOSCAN) setting; N (no) or Y
(yes).
3.
To change the setting, press
to place the cursor on the setting (N or Y) and press
Δ
or
∇
to change the setting.
4.
Press ENTER to exit from the menu structure.
NOTE
With auto scan enabled, DO NOT save the present setup as the power-on default
setup. If you do so, an interrupted scan will not resume.
If during the power-up sequence the Model 2700 detects a card ID change for
any slot, auto scan configuration will disable and an interrupted scan will not be
resumed. Error +517 occurs (cannot resume scan) to indicate that the scan has
been disabled. The instrument assumes the normal power-on setup.
The Model 7706 does not support auto scan. Trying to enable auto scan with a
Model 7706 card installed will cause error -221 (settings conflict).
Содержание 2700
Страница 3: ......
Страница 7: ......
Страница 197: ...5 24 Rel Math Ratio Channel Average dB Model 2700 Multimeter Switch System User s Manual...
Страница 213: ...6 16 Buffer Model 2700 Multimeter Switch System User s Manual...
Страница 289: ...9 18 Limits and Digital I O Model 2700 Multimeter Switch System User s Manual...
Страница 335: ...11 24 Status Structure Model 2700 Multimeter Switch System User s Manual...
Страница 336: ...12 CommonCommands...
Страница 344: ...13 SCPISignalOriented MeasurementCommands...
Страница 353: ...13 10 SCPI Signal Oriented Commands Model 2700 Multimeter Switch System User s Manual...
Страница 363: ...14 10 FORMat and Misc SYSTem Commands Model 2700 Multimeter Switch System User s Manual...
Страница 364: ...15 SCPIReferenceTables...
Страница 391: ...A 2 Specifications Model 2700 Multimeter Switch System User s Manual...
Страница 392: ...Model 2700 Multimeter Switch System User s Manual Specifications A 3...
Страница 393: ...A 4 Specifications Model 2700 Multimeter Switch System User s Manual...
Страница 394: ...Model 2700 Multimeter Switch System User s Manual Specifications A 5...
Страница 395: ...A 6 Specifications Model 2700 Multimeter Switch System User s Manual...
Страница 400: ...B Model7700ConnectionGuide...
Страница 411: ...B 12 Model 7700 Connection Guide Model 2700 Multimeter Switch System User s Manual...
Страница 412: ...C StatusandErrorMessages...
Страница 419: ...C 8 Status and Error Messages Model 2700 Multimeter Switch System User s Manual...
Страница 420: ...D SignalProcessing SequenceandDataFlow...
Страница 433: ...D 14 Signal Processing Sequence and Data Flow Model 2700 Multimeter Switch System User s Manual...
Страница 434: ...E MeasurementConsiderations...
Страница 443: ...E 10 Measurement Considerations Model 2700 Multimeter Switch System User s Manual...
Страница 454: ...G IEEE 488BusOverview...
Страница 469: ...G 16 IEEE 488 Bus Overview Model 2700 Multimeter Switch System User s Manual...
Страница 470: ...H KE2700InstrumentDriver Examples...
Страница 489: ...H 20 KE2700 Instrument Driver Examples Model 2700 Multimeter Switch System User s Manual...
Страница 499: ......