Model 2700 Multimeter/Switch System User’s Manual
Rel, Math, Ratio, Channel Average, dB
5-11
Reciprocal (1/X)
The reciprocal of a reading is displayed when the reciprocal (1/X) math function is
enabled:
Reciprocal = 1/X
where: X is the normal input reading
The displayed units designator for reciprocal readings is “R.” This units designator cannot
be changed.
Example — Assume the normal displayed reading is 2.5
Ω
. The reciprocal of resistance is
conductance. When the reciprocal math function is enabled, the following conductance
reading will be displayed:
0.4 R
Reciprocal (1/X) configuration
1.
Press SHIFT and then MATH to display the math menu.
2.
Press the RANGE
Δ
or
∇
key to display “1/X,” and press ENTER. The MATH
annunciator will turn on, and the result of the calculation will be displayed. Note
that the calculation will be applied to all measurement functions.
3.
To disable 1/X, again press SHIFT and then MATH. The MATH annunciator will
turn off.
NOTE
The result of the 1/X calculation may be displayed in exponential notation. For
example, a displayed reading of +2.500E+03 R is equivalent to 2500 R
(2.5K R).
When using Rel, the rel’ed reading of the input signal is used by the 1/X
calculation.
Содержание 2700
Страница 3: ......
Страница 7: ......
Страница 197: ...5 24 Rel Math Ratio Channel Average dB Model 2700 Multimeter Switch System User s Manual...
Страница 213: ...6 16 Buffer Model 2700 Multimeter Switch System User s Manual...
Страница 289: ...9 18 Limits and Digital I O Model 2700 Multimeter Switch System User s Manual...
Страница 335: ...11 24 Status Structure Model 2700 Multimeter Switch System User s Manual...
Страница 336: ...12 CommonCommands...
Страница 344: ...13 SCPISignalOriented MeasurementCommands...
Страница 353: ...13 10 SCPI Signal Oriented Commands Model 2700 Multimeter Switch System User s Manual...
Страница 363: ...14 10 FORMat and Misc SYSTem Commands Model 2700 Multimeter Switch System User s Manual...
Страница 364: ...15 SCPIReferenceTables...
Страница 391: ...A 2 Specifications Model 2700 Multimeter Switch System User s Manual...
Страница 392: ...Model 2700 Multimeter Switch System User s Manual Specifications A 3...
Страница 393: ...A 4 Specifications Model 2700 Multimeter Switch System User s Manual...
Страница 394: ...Model 2700 Multimeter Switch System User s Manual Specifications A 5...
Страница 395: ...A 6 Specifications Model 2700 Multimeter Switch System User s Manual...
Страница 400: ...B Model7700ConnectionGuide...
Страница 411: ...B 12 Model 7700 Connection Guide Model 2700 Multimeter Switch System User s Manual...
Страница 412: ...C StatusandErrorMessages...
Страница 419: ...C 8 Status and Error Messages Model 2700 Multimeter Switch System User s Manual...
Страница 420: ...D SignalProcessing SequenceandDataFlow...
Страница 433: ...D 14 Signal Processing Sequence and Data Flow Model 2700 Multimeter Switch System User s Manual...
Страница 434: ...E MeasurementConsiderations...
Страница 443: ...E 10 Measurement Considerations Model 2700 Multimeter Switch System User s Manual...
Страница 454: ...G IEEE 488BusOverview...
Страница 469: ...G 16 IEEE 488 Bus Overview Model 2700 Multimeter Switch System User s Manual...
Страница 470: ...H KE2700InstrumentDriver Examples...
Страница 489: ...H 20 KE2700 Instrument Driver Examples Model 2700 Multimeter Switch System User s Manual...
Страница 499: ......