Appendix B: Calibration
Model 2651A High Power System SourceMeter® Instrument Reference Manual
B-8
2651A-901-01 Rev. A / March 2011
Figure 113: Connections for current verification (1 A range and above)
LO
HI
Load
terminals
Sense
terminals
Input
LO
Model 2651A
!
LINE RATING
100-240 VAC
50, 60 Hz
500VA MAX.
LINE FUSE
SLOWBLOW
5.0 A, 250 V
NO INTERNAL OPERATOR SERVICEABLE PARTS,
SERVICE BY QUALIFIED PERSONNEL ONLY.
REPLACE FUSE WITH SAME TYPE AND RATING.
WARNING:
CHANNEL A
RS-232
DIGITAL I/O
LAN
!
IEEE-488
G
G G G
S
LO
SENSE/GUARD
S
HI
LO
HI
OUTPUT
40V, 50A MAX.
!
Input
HI
Characterized
resistor
Digital multimeter
0.1
Ω
250 W
Current measurement accuracy
Follow the steps below to verify that Model 2651A current measurement accuracy is within specified
limits. The procedure involves applying accurate currents from the Model 2651A current source and
then verifying that Model 2651A current measurements are within required limits.
1. With the power off, connect the digital multimeter to the Model 2651A terminals as shown in
Connections for current verification (100 mA range and below)
2. Select the multimeter DC current function.
3. Set the Model 2651A to both source and measure current by pressing the
SRC
and then the
MEAS
keys. Make sure the source output is turned on.