![Keithley 2651A Скачать руководство пользователя страница 689](http://html1.mh-extra.com/html/keithley/2651a/2651a_reference-manual_661729689.webp)
Section 8: Troubleshooting guide
Model 2651A High Power System SourceMeter® Instrument Reference Manual
8-6
2651A-901-01 Rev. A / March 2011
Error summary
Error number
Error level
Error Message
5044
SERIOUS
Cannot perform requested operation while current measure
autorange is enabled
5045
SERIOUS
Cannot perform requested operation while filter is enabled
5046
SERIOUS
SMU too hot
5047
RECOVERABLE
Minimum timestamp resolution is 1
μ
s
5048
SERIOUS
Contact check not valid with HIGH-Z output off
5049
SERIOUS
Contact check not valid while an active current source
5050
SERIOUS
I limit too low for contact check
5051
FATAL
Model number/SMU hardware mismatch. Disconnect DUT and cycle
power
5052
RECOVERABLE
Interlock engaged; system stabilizing
5053
RECOVERABLE
Unstable output detected - Measurements may not be valid
5054
RECOVERABLE
High C voltage limit exceeded
5055
SERIOUS
Cannot change adjustment date - change cal constants before
operation
5059
SERIOUS
trigger.source.action enabled without configuration
5060 SERIOUS trigger.measure.action enabled without configuration
5061
SERIOUS
Operation not permitted while OUTPUT is off
5063
SERIOUS
Cannot perform requested operation while measure autozero is on
5064
SERIOUS
Cannot use reading buffer that collects source values
5065
SERIOUS
I range too low for contact check
5066
SERIOUS
source.offlimiti too low for contact check
5069
SERIOUS
Autorange locked for HighC mode
LAN troubleshooting suggestions
If you are unable to connect to the instrument's web interface, check the following items:
•
Verify that the network cable is in the LAN port, not one of the TSP-Link
®
ports (see the
(on page 2-6)).
•
Verify that the network cable is in the correct port on the computer. The side LAN port of a laptop
may be disabled while the unit is in a docking station.
•
Verify that the correct Ethernet card's configuration information was used during the setup
procedure.
•
Verify that the computer's network card is enabled.
•
Verify the instrument's IP address is compatible with the IP address on the computer.
•
Verify the instrument's subnet mask address is the same as the computer's subnet mask
address.
•
Turn the instrument's power off, and then on.
•
Restart your computer.
•
Contact your system administrator for assistance.