![Keithley 2651A Скачать руководство пользователя страница 591](http://html1.mh-extra.com/html/keithley/2651a/2651a_reference-manual_661729591.webp)
Model 2651A High Power System SourceMeter® Instrument Reference Manual
Section 7: Command reference
2651A-901-01 Rev. A / March 2011
7-283
status.operation.instrument.trigger_blender.*
This attribute contains the operation status trigger blender summary register set.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute
- -
- -
- -
- -
.condition (R)
Yes
Not applicable
Not saved
Not applicable
.enable (RW)
Yes
Status reset
Not saved
0
.event (R)
Yes
Status reset
Not saved
0
.ntr (RW)
Yes
Status reset
Not saved
0
.ptr (RW)
Yes
Status reset
Not saved
1024 (All bits set)
Usage
operationRegister
= status.operation.instrument.trigger_blender.condition
operationRegister
= status.operation.instrument.trigger_blender.enable
operationRegister
= status.operation.instrument.trigger_blender.event
operationRegister
= status.operation.instrument.trigger_blender.ntr
operationRegister
= status.operation.instrument.trigger_blender.ptr
status.operation.instrument.trigger_blender.enable =
operationRegister
status.operation.instrument.trigger_blender.ntr =
operationRegister
status.operation.instrument.trigger_blender.ptr =
operationRegister
operationRegister
The operation status trigger blender summary register's status. A zero (0) indicates
no bits set (also send 0 to clear all bits); the only valid value other than 0 is 1024
Details
These attributes are used to read or write to the operation status trigger blender summary registers. Reading a
status register returns a value. The binary equivalent of the returned value indicates which register bits are set.
The least significant bit of the binary number is Bit B0, and the most significant bit is Bit B15.
For information about .condition, .enable, .event, .ntr, and .ptr registers, refer to
(on page E-
Enable and transition registers
(on page E-18). The individual bits of this register have the following
meanings:
Bit Value
Description
B0-B9 Not
used
Not applicable.
B10
status.operation.instrument.trigger_blender.TRIGGER_OVERRUN
status.operation.instrument.trigger_blender.TRGOVR
Set bit indicates one
or more enabled bits
for operation status
trigger blender
overrun register is
set.
Bit B10 decimal
value: 1024
Binary value:
0100 0000 0000
B11-
B15
Not used
Not applicable.