Section 7: Command reference
Model 2651A High Power System SourceMeter® Instrument Reference Manual
7-104
2651A-901-01 Rev. A / March 2011
Details
Disconnects all existing LAN connections to the instrument and re-initializes the LAN with the current
configuration settings.
This function initiates a background operation. LAN configuration could be a lengthy operation. Although the
function returns immediately, the LAN initialization will continue to run in the background.
Even though the LAN configuration settings may not have changed since the LAN was last connected, new
settings may take effect due to the dynamic nature of DHCP or DLLA configuration.
Re-initialization takes effect even if the configuration has not changed since the last time the instrument
connected to the LAN.
Example
lan.applysettings()
Re-initialize the LAN interface with new settings.
Also see
None
lan.autoconnect
This attribute is used to enable or disable link monitoring.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
LAN restore defaults
Nonvolatile memory lan.ENABLE
Usage
state
= lan.autoconnect
lan.autoconnect =
state
state
LAN link monitoring state:
1
or
lan.ENABLE:
Enables automatic link reconnection and monitoring
0
or
lan.DISABLE:
Disables automatic link reconnection and monitoring
Details
This attribute sets the LAN link monitoring and automatic connection state.
When this is set to
lan.ENABLE
, all connections are closed if the link to the LAN is lost for more than the time
specified by
lan.linktimeout
.
Set this attribute to
lan.ENABLE
to automatically reset the LAN configuration after the LAN link is established.
Also see
(on page 7-112)
lan.config.dns.address[N]
Configures DNS server IP addresses.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
LAN restore defaults
Nonvolatile memory "0.0.0.0"