
Section 9: Frequently asked questions
2606B System SourceMeter® Instrument Reference Manual
9-4
2606B-901-01 Rev. B / May 2018
How do I store measurements in nonvolatile memory?
After the measurements are complete, you can save the reading buffer data to the nonvolatile
memory in the instrument.
To save the reading buffer data:
1. From the virtual front panel, press the
STORE
key, and then select
SAVE
.
2. Select
INTERNAL
to save to internal nonvolatile memory.
3. Select one of the following:
SMUA_BUFFER1
SMUA_BUFFER2
SMUB_BUFFER1
SMUB_BUFFER2
4. The virtual front panel displays
Saving... This may take awhile
.
5. Press the
EXIT (LOCAL)
key to return to the main menu.
For additional information, see
(on page 3-8).
When should I change the output-off state?
Carefully consider and configure the appropriate output-off state, source, and compliance
limits before connecting the 2606B to a device that can deliver energy (for example, other
voltage sources, batteries, capacitors, solar cells, or other 2606B instruments). Configure
recommended instrument settings before making connections to the device. Failure to
consider the output-off state, source, and compliance limits may result in damage to the
instrument or to the device under test (DUT).
The 2606B instrument provides multiple output-off states. The multiple states are required because
different types of connected devices (or loads) require different behaviors from the 2606B when its
output is turned off.
For example, a passive device such as a diode is not affected by a 0 V source connected across its
terminals when the output is turned off. However, connecting a 0 V source to the terminals of a
battery causes the battery to discharge. Therefore, careful selection of the proper output-off state is
important to prevent damage to devices and instruments. This is especially true when the device can
deliver energy to the 2606B, such as a battery or capacitor or when another SourceMeter instrument
is connected across the output terminals. In these situations, you should use an output-off state that
isolates the instrument from the device by either setting
smu
X
.source.offfunc =
smu
X
.OUTPUT_DCAMPS
or
smu
X
.source.offfunc = smu
X
.OUTPUT_DCVOLTS
, as applicable.
There are other guidelines to follow when connecting the output of multiple 2606B instruments to get
a larger current or voltage. Please refer to the following references for more information:
•
Combining module outputs
•
Keithley application notes on