2606B System SourceMeter® Instrument Reference Manual
Section 7: TSP command reference
2606B-901-01 Rev. B / May 2018
7-359
tsplink.state
This attribute describes the TSP-Link online state.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (R)
Yes
Not applicable
Not applicable
Not applicable
Usage
state
= tsplink.state
state
TSP-Link state (
online
or
offline
)
Details
When the instrument power is first turned on, the state is
offline
. After
tsplink.reset()
is
successful, the state is
online
.
Example
state = tsplink.state
print(state)
Read the state of the TSP-Link system. If it is online, the output is:
online
Also see
(on page 7-356)
(on page 7-358)
tsplink.trigger[N].assert()
This function simulates the occurrence of the trigger and generates the corresponding event ID.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
tsplink.trigger[
N
].assert()
N
The trigger line (1 to 3)
Details
The set pulse width determines how long the trigger is asserted.
Example
tsplink.trigger[2].assert()
Asserts trigger on trigger line 2.
Also see
(on page 7-360)
(on page 7-361)
(on page 7-362)
(on page 7-363)
(on page 7-363)
(on page 7-365)
(on page 7-366)