Section 2: General operation
2606B System SourceMeter® Instrument Reference Manual
2-32
2606B-901-01 Rev. B / May 2018
Press the
EXIT
key to return to a previous menu.
The following table contains descriptions of the configuration menus with cross references to related
information.
To select a menu:
1. Select the
CONFIG
key.
2. Select
CHANNEL-A
,
CHANNEL-B
, or
COMMON
. COMMON accesses settings for TRIG and
STORE.
3. Select the key associated with the menu.
To access, select:
Options
For more information, see:
SRC
V-source sense, low range; I-source
low range; and highC mode
(on page 2-71),
(on page 2-39)
MEAS
V and I-measure range, V-measure
sense, low range; autozero
(on page 2-71),
(on page 2-39)
LIMIT
V-source and I-source compliance
limits
(on page 2-35)
SPEED
Measurement speed (NPLC)
(on page 2-76)
REL
Relative values
(on page 3-1)
FILTER
Control digital filter
(on page 3-3)
OUTPUT ON/OFF
Off-state, control digital I/O
(on page 2-65)
TRIG
Trigger in, count, interval, and delay
(on page 3-30)
STORE
Buffer count and destination
(on page 4-
Setting a value
Through the virtual front panel, you can adjust a value using the Navigation wheel and the keypad.
To adjust a value:
1. Use the
CURSOR
arrow keys or the navigation wheel to move the cursor to the digit that needs
to be changed.
2. Select the navigation wheel or the
ENTER
key to enter edit mode. The EDIT indicator is
illuminated.
3. Rotate the navigation wheel or use the number pad to set the value. If you use the number pad,
the cursor moves to the next digit on the right.
4. Select the
ENTER
key to select the value or press the
EXIT (LOCAL)
key to cancel the change.
5. To return to the main menu, press the
EXIT (LOCAL)
key.
To set an option to zero (if it is an option for the entry), select the
0000
numeric entry key. To toggle
the polarity of a value, press the
+/–
numeric entry key.