2606B System SourceMeter® Instrument Reference Manual
Section 7: TSP command reference
2606B-901-01 Rev. B / May 2018
7-227
Example
smua.trigger.arm.set()
Sets the arm event detector to the detected state for SMU channel A.
Also see
(on page 7-236)
(on page 7-246)
smuX.trigger.arm.stimulus
This attribute selects the event that will cause the arm event detector to enter the detected state.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
SMU reset
Instrument reset
Recall setup
Not saved
0
Usage
eventID
= smu
X
.trigger.arm.stimulus
smu
X
.trigger.arm.stimulus =
eventID
eventID
Event that triggers the arm detector
X
Source-measure unit (SMU) channel (for example,
smua.trigger.arm.stimulus
applies to SMU channel A)
Details
Set this attribute to the event ID of any trigger event generator to wait for that event.
Set this attribute to zero to bypass waiting for events at the arm event detector (the SMU continues
uninterrupted through the remote trigger model). Set
eventID
to one of the existing trigger event IDs
shown in the following table.