2606B System SourceMeter® Instrument Reference Manual
Section 7: TSP command reference
2606B-901-01 Rev. B / May 2018
7-195
Also see
(on page B-10)
(on page 7-180)
(on page 7-182)
(on page 7-183)
(on page 7-213)
smuX.measure.count
This attribute sets the number of measurements made when a measurement is requested.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Instrument reset
SMU reset
Recall setup
Saved setup
1
Usage
count
= smu
X
.measure.count
smu
X
.measure.count =
count
count
Number of measurements
X
Source-measure unit (SMU) channel (for example,
smua.measure.count
applies
to SMU channel A)
Details
This attribute controls the number of measurements made any time a measurement is requested.
When using a reading buffer with a measure command, this attribute also controls the number of
readings to be stored.
If
smu
X
.measure.count
is set to a value greater than 1, any measurement delay set by
smu
X
.measure.delay
only occurs before the first measurement, while the
smu
X
.measure.interval
controls the interval between successive measurements.
Example
smua.measure.count = 10
Sets the SMU channel A measure count to 10.
Also see
(on page 7-159)
(on page 7-173)
(on page 7-196)
(on page 7-200)
(on page 7-210)